HP manual HP 8753ET specifications, Test-port output characteristics, KHz to 3 GHz

Page 14

HP 8753ET specifications

Test-port output characteristics

Frequency characteristics

Range

300 kHz to 3 GHz

Option 006 300 kHz to 6 GHz

Resolution

1 Hz

Stability4

±7.5 ppm (0˚ to 55 ˚C)

 

±3 ppm/year

Option 1D5 ±0.05 ppm (0˚ to 55 ˚C)

 

±0.5 ppm/year

Accuracy

±10 ppm at 25 ˚C ±5 ˚C

Power range:

-20 to +5 dBm

 

-85 to +10 dBm (with Option 004)

Resolution

0.01 dB

Level accuracy1

±1.0 dB

Level linearity1

 

Standard

 

-20 to -15 dBm

±0.5 dB

-15 to 0 dBm

±0.2 dB

0 to +5 dBm

±0.5 dB

Option 004

 

-15 to +5 dBm

±0.2 dB

+5 to +10 dBm

±0.5 dB

Impedance

50 Ω (nominal)

300 kHz to 3 GHz >18 dB RL (< 1.28 SWR)

3 GHz to 6 GHz

> 14 dB RL (<1.50 SWR)

Spectral purity

 

2nd harmonic2

 

 

<-40 dBc at 0 dBm4

 

<-50 dBc at -10 dBm4

Option 002

<-25 dBc at maximum output

 

power

3rd harmonic3

 

 

<-40 dBc at 0 dBm

 

<-50 dBc at -10 dBm

Option 002

<-25 dBc at maximum output

 

power

Nonharmonic spurious

Mixer related

<-30 dBc at +10 dBm

 

<-55 dBc at -10 dBm

Test port input characteristics

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

Average noise level

 

 

300 kHz to 3 GHz

<-90 dBm (3 kHz BW)

300 kHz to 3 GHz

<-110 dBm (10 Hz BW)

300 kHz to 3 GHz

<-120 dBm (10 Hz BW) typical

3 GHz to 6 GHz

<-85 dBm (3 kHz BW)

3 GHz to 6 GHz

<-105 dBm (10 Hz BW)

3 GHz to 6 GHz

<-114 dBm (10 Hz BW) typical

Maximum input level

+10 dBm reflection port

 

0 dBm transmission port

Damage level

>+26 dBm or 35 VDC

Impedance

50 Ω (nominal)

Frequency response5

 

 

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)

 

2nd harmonic2

<-15 dBc at +8 dBm

 

<-30 dBc at 0 dBm4

 

<-45 dBc at -15 dBm4

3rd harmonic3

<-30 dBc at +8 dBm

 

<-50 dBc at 0 dBm4

 

<-50 dBc at -15 dBm4

Harmonic measurement accuracy5

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range4

-40 dBc

(output = -10 dBm,

Frequency offset mode6

input = <-15 dBm)

 

 

Frequency range

 

300 kHz to 3 GHz

Option 006

 

300 kHz to 6 GHz

R-channel input requirements

Power level

 

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

 

0 to -30 dBm

LO Spectral purity

 

 

Maximum spurious

<-25 dBc

input

 

 

Residual FM

 

<20 kHz

LO frequency accuracy6

±1 MHz at nominal

 

 

frequency

1.Relative to -5 dBm output power for the HP 8753ET; -10 dBm output power for HP 8753ET with Option 004.

2.16 MHz to 3 GHz.

3.16 MHz to 2 GHz.

4.Typical performance.

5.25˚ C ±5˚ C.

6.The HP 8753ET source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.

14

Image 14
Contents Specifications Table of contents Dynamic range Measurement port characteristicsDefinitions and test conditions Measurement uncertaintyMm test ports System performance summariesHP 8753ES 50-ohm systems Network analyzer HP 8753ESType-N test ports Calibration kit HP 85032B Test-port cables HP 11857DCalibration kit HP 85033D Test-port cables HP 11857D Frequency rangeCalibration kit HP 85036B Test-port cables HP 11857B HP 8753ES 75-ohm systemsNetwork analyzer Type-F test ports Calibration kit HP 85039B Test-port cables HP 11857BMeasurement port characteristics KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dBTest port input characteristics HP 8753ES specificationsTest-port output characteristics7 CW time sweep only Characteristics Phase characteristics KHz to 6 GHzTest port input characteristics Option HP 8753ES Option 011 specificationsTest port output characteristics Frequency KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm20 kHz CW time sweep onlyKHz to 3 GHz HP 8753ET specificationsTest-port output characteristics Option 006 300 kHz to 6 GHz±500 dB Test-port input characteristicsMagnitude characteristics ±180˚Measurement HP 8753ET/ES supplemental characteristicsGroup delay characteristics Source controlDisplay control Data accuracy enhancementData averaging Data hardcopy Trace functionsStorage System capabilitiesTime-domain Option Remote programmingTest sequences This stimulus is also used to measure low-pass devicesGeneral characteristics Options HP 85046A/B S-parameter test sets Test set specifications for OptionHP 85046A/B Specifications +20 dBm SpecificationsHP 85047A S-parameter test set +30 dBmAccessories HP 11906B 7-16 to 50-Ohm type-N adapter kit HP 85039B type-F calibration kit SpecificationsHP 85038 family of calibration kits HP 11906C 7-16 to 7-mm adapter kitVerification kits Test-port return cablesHP 85024A high-frequency probe Ohm accessory kitsRF limiter HP 8347A RF amplifier DimensionsWeight