HP 8753ET KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm, 20 kHz, Frequency, CW time sweep only

Page 13

HP 8753ES Option 011 specifications

Test port input characteristics (continued)

Frequency offset mode3

 

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

Power level

 

300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

LO spectral purity

 

Maximum spurious

<-25 dBc

input

 

Residual FM

<20 kHz

LO frequency accuracy ±1 MHz at nominal

 

frequency

External source mode4

(CW time sweep only)

Frequency range

300 kHz to 6 GHz

R-channel input requirements1

Power level

0 to -25 dBm

Spectral purity

 

Maximum spurious

<-30 dBc

input

 

Residual FM

<20 kHz

Typical settling time1

500 ms (automatic)

 

50 ms (manual)

Input frequency accuracy requirement1

Manual:

-0.5 to 5 MHz

Display resolution

0.001 dB/division

Marker resolution5

0.001 dB

Trace noise

 

(+5 dBm at test-port, ratio measurement, 3 kHz BW)

Magnitude

 

300 kHz to 3 GHz

< 0.006 dB rms

3 GHz to 6 GHz

< 0.010 dB rms

Phase

 

300 kHz to 3 GHz

< 0.038˚ rms

3 GHz to 6 GHz

< 0.070˚ rms

Reference level

 

Range

±500 dB

Resolution

0.001 dB

Stability2

 

30 kHz to 3 GHz

0.02 dB/ ˚C

3 GHz to 6 GHz

0.04 dB/ ˚C

Range

±180˚

Display resolution

0.01˚/ division

Marker resolution5

0.01˚

Trace noise2

 

30 kHz to 3 GHz

< 0.038˚ rms

3 GHz to 6 GHz

< 0.070˚ rms

 

(+5 dBm at test port, ratio

 

measurement, 3 kHz BW)

Reference level

 

Range

-180˚ to +180˚

Resolution

0.01˚

Stability

 

30 kHz to 3 GHz

0.05˚/ ˚C

3 GHz to 6 GHz

0.20˚/ ˚C

Polar characteristics

 

Range

10 x 10-12to 1000 units full

 

scale

Reference

±500 units

1.Typical performance.

2.Typical below 300 kHz.

3.The HP 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.

4.See the HP 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal.

5.Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.

13

Image 13
Contents Specifications Table of contents Definitions and test conditions Measurement port characteristicsDynamic range Measurement uncertaintyHP 8753ES 50-ohm systems System performance summariesMm test ports Network analyzer HP 8753ESCalibration kit HP 85032B Test-port cables HP 11857D Type-N test portsFrequency range Calibration kit HP 85033D Test-port cables HP 11857DNetwork analyzer HP 8753ES 75-ohm systemsCalibration kit HP 85036B Test-port cables HP 11857B Calibration kit HP 85039B Test-port cables HP 11857B Type-F test portsKHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dB Measurement port characteristicsTest-port output characteristics7 HP 8753ES specificationsTest port input characteristics Phase characteristics CharacteristicsCW time sweep only KHz to 6 GHzTest port output characteristics HP 8753ES Option 011 specificationsTest port input characteristics Option 20 kHz KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBmFrequency CW time sweep onlyTest-port output characteristics HP 8753ET specificationsKHz to 3 GHz Option 006 300 kHz to 6 GHzMagnitude characteristics Test-port input characteristics±500 dB ±180˚Group delay characteristics HP 8753ET/ES supplemental characteristicsMeasurement Source controlData averaging Data accuracy enhancementDisplay control Storage Trace functionsData hardcopy System capabilitiesTest sequences Remote programmingTime-domain Option This stimulus is also used to measure low-pass devicesGeneral characteristics Options HP 85046A/B Specifications Test set specifications for OptionHP 85046A/B S-parameter test sets HP 85047A S-parameter test set Specifications+20 dBm +30 dBmAccessories HP 85038 family of calibration kits HP 85039B type-F calibration kit SpecificationsHP 11906B 7-16 to 50-Ohm type-N adapter kit HP 11906C 7-16 to 7-mm adapter kitTest-port return cables Verification kitsRF limiter Ohm accessory kitsHP 85024A high-frequency probe Weight DimensionsHP 8347A RF amplifier

8753ET specifications

The HP 8753ET is a sophisticated vector network analyzer designed to meet the demands of RF and microwave engineering applications. This instrument is renowned for its ability to provide high-performance measurements in a compact form factor, making it a favorite among engineers and technicians alike.

One of the main features of the HP 8753ET is its ability to perform a wide array of network analysis functions. It supports S-parameter measurements, allowing for the characterization of the linearity and performance of RF circuits. With the ability to measure reflection, transmission, and impedance characteristics, users can gain deep insights into their device under test (DUT).

An essential aspect of the HP 8753ET is its frequency range capability. It typically covers frequencies from 300 kHz to 6 GHz, which is ideal for a vast range of applications, including telecommunications, radar, and electronic warfare.

The HP 8753ET is equipped with advanced analysis features such as time-domain reflectometry (TDR) and gating capabilities. This enables engineers to identify and analyze impedance mismatches and faults over a defined time window, helping to troubleshoot and optimize RF systems effectively.

Another notable technology integrated into the HP 8753ET is its user-friendly interface that combines an easy-to-navigate display with direct access to measurement functions. This minimizes the learning curve for new users while enhancing productivity for experienced professionals.

The analyzer also supports various calibration techniques, ensuring accurate and repeatable measurements. The built-in calibration kits help improve the reliability of the results while compensating for any systematic errors associated with the test setup.

Additional characteristics of the HP 8753ET include its capability for both single-port and two-port measurements. This versatility allows users to conduct a complete array of S-parameter measurements with confidence.

Furthermore, the HP 8753ET features a storage function that allows users to save and recall measurement setups and results. This capability streamlines workflows and provides easy access to historical data for review or further analysis.

Ultimately, the HP 8753ET stands out as an essential tool for engineers in the RF and microwave domain. Its combination of high-performance measurement capabilities, an intuitive interface, and flexibility makes it a preferred choice for anyone seeking reliable and thorough network analysis in their testing and development projects.