HP 8753ET manual Test set specifications for Option, HP 85046A/B S-parameter test sets

Page 22

Test set specifications for Option 011

HP 85046A/B S-parameter test sets

The HP 85046A/B S-parameter test sets provide the capability to measure reflection and transmission char- acteristics (including S-parameters) of two-port devices in either direction with a single connection. The test sets are controlled from the HP 8753ES Option 011 and include a programmable step attenua- tor. The frequency range of the HP 85046A 50-ohm test set is 300 kHz to 3 GHz. The HP 85046A has preci- sion 7-mm connectors. The frequency range of the HP 85046B 75-ohm test set is 300 kHz to 2 GHz. The HP 85046B has 75-ohm type-N(f) connectors. Both connectors can be adapted to other interfaces with the appropriate precision adapters.

HP 85046A/B Specifications

Note: Specifications that apply only to the HP 85046 are indicated in parentheses.

Impedance

50 ohm (75 ohm)

Frequency range

300 kHz to 3 GHz

 

(300 kHz to 2 GHz)

Directivity

35 dB to 1.3 GHz

Typical tracking

30 dB to Fmax1

 

Transmission magnitude, phase2

0.3

MHz to 2.0 MHz

±1.5 dB, ±20˚

2.0

MHz to Fmax

±1.5 dB, ±10˚

Reflection magnitude, phase2

0.3

MHz to 2.0 MHz

±1.5 dB, ±25˚

2.0

MHz to Fmax

±1.5 dB, ±10˚

Effective source match

0.3

MHz to 2.0 MHz

14 dB

2.0

MHz to 1.3 GHz

20 dB (17 dB)

1.3

GHz to Fmax

16 dB

Nominal insertion loss

14 dB + 0.5 dB/GHz

Input to test port

 

 

(19.5 dB + 1 dB/GHz)

Input to incident

18 dB + 1.5 dB/GHz

 

 

(18 dB + 1.5 dB/GHz)

Port 1, 2 to A, B

6.5 dB + 1.0 dB/GHz

 

 

(12 dB + 0.5 dB GHz)

Test set switch/repeatability

 

 

±0.03 dB; ±0.01 dB3

Max. operating level

+20 dBm

Damage level

+30 dBm

RF attenuator range

70 dB, 10 dB steps

DC bias range

±30 VDC, 200 mA4

 

 

500 mA max

DC bias connectors

50 ohm BNC (f)

Includes

four 190 mm (7.5 in) type-N

 

 

cables and test set

 

 

interconnect cable.

Dimensions

90 mm H x 432 mm W x 553 mm D

Weight

 

9.1 kg (20 lb)

A standard HP 85046A/B test set contains a solid-state transfer switch, which allows continuous switching of power from port 1 to port 2 for full two-port error cor- rection. Option 009 replaces the transfer switch with a mechanical switch. This provides about 1.5 dB more power at the test port, but does not allow continuous switching, so the user must initiate updates of all four S-parameters for full two-port error correction. Also, the mechanical switch has relays that will wear out faster than the solid-state switch. Approximate lifetime of the mechanical switch is 1 million cycles.

HP 85046A schematic

1.Fmax is the upper frequency limit of the associated test set.

2.Degrees, specified as deviation from linear phase.

3.Typical performance.

4. Some degradation of RF specifications may occur.

 

22

 

 

Image 22
Contents Specifications Table of contents Dynamic range Measurement port characteristicsDefinitions and test conditions Measurement uncertaintyMm test ports System performance summariesHP 8753ES 50-ohm systems Network analyzer HP 8753ESType-N test ports Calibration kit HP 85032B Test-port cables HP 11857DCalibration kit HP 85033D Test-port cables HP 11857D Frequency rangeNetwork analyzer HP 8753ES 75-ohm systemsCalibration kit HP 85036B Test-port cables HP 11857B Type-F test ports Calibration kit HP 85039B Test-port cables HP 11857BMeasurement port characteristics KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dBTest-port output characteristics7 HP 8753ES specificationsTest port input characteristics CW time sweep only CharacteristicsPhase characteristics KHz to 6 GHzTest port output characteristics HP 8753ES Option 011 specificationsTest port input characteristics Option Frequency KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm20 kHz CW time sweep onlyKHz to 3 GHz HP 8753ET specificationsTest-port output characteristics Option 006 300 kHz to 6 GHz±500 dB Test-port input characteristicsMagnitude characteristics ±180˚Measurement HP 8753ET/ES supplemental characteristicsGroup delay characteristics Source controlData averaging Data accuracy enhancementDisplay control Data hardcopy Trace functionsStorage System capabilitiesTime-domain Option Remote programmingTest sequences This stimulus is also used to measure low-pass devicesGeneral characteristics Options HP 85046A/B Specifications Test set specifications for OptionHP 85046A/B S-parameter test sets +20 dBm SpecificationsHP 85047A S-parameter test set +30 dBmAccessories HP 11906B 7-16 to 50-Ohm type-N adapter kit HP 85039B type-F calibration kit SpecificationsHP 85038 family of calibration kits HP 11906C 7-16 to 7-mm adapter kitVerification kits Test-port return cablesRF limiter Ohm accessory kitsHP 85024A high-frequency probe Weight DimensionsHP 8347A RF amplifier

8753ET specifications

The HP 8753ET is a sophisticated vector network analyzer designed to meet the demands of RF and microwave engineering applications. This instrument is renowned for its ability to provide high-performance measurements in a compact form factor, making it a favorite among engineers and technicians alike.

One of the main features of the HP 8753ET is its ability to perform a wide array of network analysis functions. It supports S-parameter measurements, allowing for the characterization of the linearity and performance of RF circuits. With the ability to measure reflection, transmission, and impedance characteristics, users can gain deep insights into their device under test (DUT).

An essential aspect of the HP 8753ET is its frequency range capability. It typically covers frequencies from 300 kHz to 6 GHz, which is ideal for a vast range of applications, including telecommunications, radar, and electronic warfare.

The HP 8753ET is equipped with advanced analysis features such as time-domain reflectometry (TDR) and gating capabilities. This enables engineers to identify and analyze impedance mismatches and faults over a defined time window, helping to troubleshoot and optimize RF systems effectively.

Another notable technology integrated into the HP 8753ET is its user-friendly interface that combines an easy-to-navigate display with direct access to measurement functions. This minimizes the learning curve for new users while enhancing productivity for experienced professionals.

The analyzer also supports various calibration techniques, ensuring accurate and repeatable measurements. The built-in calibration kits help improve the reliability of the results while compensating for any systematic errors associated with the test setup.

Additional characteristics of the HP 8753ET include its capability for both single-port and two-port measurements. This versatility allows users to conduct a complete array of S-parameter measurements with confidence.

Furthermore, the HP 8753ET features a storage function that allows users to save and recall measurement setups and results. This capability streamlines workflows and provides easy access to historical data for review or further analysis.

Ultimately, the HP 8753ET stands out as an essential tool for engineers in the RF and microwave domain. Its combination of high-performance measurement capabilities, an intuitive interface, and flexibility makes it a preferred choice for anyone seeking reliable and thorough network analysis in their testing and development projects.