Environmental characteristics

Measurement throughput summary

General conditions

RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards.

ESD (electrostatic discharge): must be eliminated by use of static-safe work procedures and an anti-static bench mat. The flexible rubber keypad protects key contacts from dust, but the environment should be as dust-free as possible for optimal reliability.

Operating conditions

 

Temperature

0˚ to 55 ˚C

(unless otherwise noted)

Humidity

5% to 95% at 40 ˚C

 

(non-condensing)

Altitude

0 to 4500 meters

 

(15,000 feet)

Non-operating storage conditions

Temperature

-40 ˚C to +70 ˚C

Humidity

0 to 90% relative at

 

+65 ˚C (non-

 

condensing)

Altitude

0 to 15,240 meters

 

(50,000 feet)

Weight (HP 8753ET/ES)

21 kg (46 lb)

Net

Shipping

35 kg (77 lb)

Cabinet dimensions (HP 8753ET/ES) (These dimensions exclude front and rear panel protrusions.)

222 mm H x 425 mm W x 457 mm D (8.75 in x 16.75 in x 18.0 in)

HP 8753ET/ES physical dimensions

The following table shows typical measurement times in milliseconds.

Typical time for completion (msec)

Number of Points

Measurement

51

201

401

1601

Start=1 GHz, Span=10 MHz, IFBW=6 kHz

 

 

 

 

Uncorrected, 1-port calibration:1

32

70

121

423

 

Two-port calibration:2

62

139

240

848

Start=30 kHz, Stop=3 GHz, IFBW=6 kHz

 

 

 

 

Uncorrected, 1-port calibration1

202

270

304

615

 

Two-port calibration2

402

540

607

1237

Start=30 kHz, Stop=6 GHz, IFBW=6 kHz

 

 

 

 

Uncorrected, 1-port calibration1

310

380

415

658

 

Two-port calibration2

618

757

829

1315

 

 

 

Time domain (increase over uncorrected sweep time)3

 

Conversion

12

42

86

378

Gating

14

40

80

349

GP-IB data transfer4

 

 

 

 

 

Internal binary

10

16

21

58

 

ASCII

35

112

214

831

 

IEEE 754 floating point format

 

 

 

 

 

32-bit

11

19

28

83

 

64-bit

13

26

42

141

Options

Harmonic measurements (Option 002)

Measures amplifier 2nd and 3rd harmonics on a swept- frequency basis for fundamental signals above 16 MHz.

Dynamic range (source at +10 dBm,

receiver <+30 dBm)

+40 dBc (minimum)

Accuracy5

±1 dB (< 6 GHz)

Step attenuator (Option 004) Provides source output power range from -85 to +10 dBm for HP 8753ET model.

6 GHz operation (Option 006)

With Option 006, performance is extended to 6 GHz. When external source, tuned receiver or harmonic mode is used, the receiver is capable of measuring sig- nals up to 6 GHz.

Time domain (Option 010)

Transforms data from the frequency domain to the time domain using a Fourier transformation technique.

High-stability frequency reference (Option 1D5) This option adds an ovenized 10-MHz frequency refer- ence output. It is connected to the external reference input on the rear panel. See the “General Characteristics” section for below specifications.

1.One-port calibration, with a 6 kHz IF bandwidth. Includes system retrace time, but does not include bandswitch time. Time-domain gating is assumed off.

2.Same as footnote 1, but for an S21 measurement with full two-port calibration. Includes RF switching time.

3.Option 010 only, gating off.

4.Measured with an HP Omnibook 7100 266 MHz Pentium® II computer.

5. Does not include error from the HP 8753T/ES source and receiver harmonics.

21

 

 

Image 21
Contents Specifications Table of contents Definitions and test conditions Measurement port characteristicsDynamic range Measurement uncertaintyHP 8753ES 50-ohm systems System performance summariesMm test ports Network analyzer HP 8753ESCalibration kit HP 85032B Test-port cables HP 11857D Type-N test portsFrequency range Calibration kit HP 85033D Test-port cables HP 11857DHP 8753ES 75-ohm systems Network analyzerCalibration kit HP 85036B Test-port cables HP 11857B Calibration kit HP 85039B Test-port cables HP 11857B Type-F test portsKHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dB Measurement port characteristicsHP 8753ES specifications Test-port output characteristics7Test port input characteristics Phase characteristics CharacteristicsCW time sweep only KHz to 6 GHzHP 8753ES Option 011 specifications Test port output characteristicsTest port input characteristics Option 20 kHz KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBmFrequency CW time sweep onlyTest-port output characteristics HP 8753ET specificationsKHz to 3 GHz Option 006 300 kHz to 6 GHzMagnitude characteristics Test-port input characteristics±500 dB ±180˚Group delay characteristics HP 8753ET/ES supplemental characteristicsMeasurement Source controlData accuracy enhancement Data averagingDisplay control Storage Trace functionsData hardcopy System capabilitiesTest sequences Remote programmingTime-domain Option This stimulus is also used to measure low-pass devicesGeneral characteristics Options Test set specifications for Option HP 85046A/B SpecificationsHP 85046A/B S-parameter test sets HP 85047A S-parameter test set Specifications+20 dBm +30 dBmAccessories HP 85038 family of calibration kits HP 85039B type-F calibration kit SpecificationsHP 11906B 7-16 to 50-Ohm type-N adapter kit HP 11906C 7-16 to 7-mm adapter kitTest-port return cables Verification kitsOhm accessory kits RF limiterHP 85024A high-frequency probe Dimensions WeightHP 8347A RF amplifier

8753ET specifications

The HP 8753ET is a sophisticated vector network analyzer designed to meet the demands of RF and microwave engineering applications. This instrument is renowned for its ability to provide high-performance measurements in a compact form factor, making it a favorite among engineers and technicians alike.

One of the main features of the HP 8753ET is its ability to perform a wide array of network analysis functions. It supports S-parameter measurements, allowing for the characterization of the linearity and performance of RF circuits. With the ability to measure reflection, transmission, and impedance characteristics, users can gain deep insights into their device under test (DUT).

An essential aspect of the HP 8753ET is its frequency range capability. It typically covers frequencies from 300 kHz to 6 GHz, which is ideal for a vast range of applications, including telecommunications, radar, and electronic warfare.

The HP 8753ET is equipped with advanced analysis features such as time-domain reflectometry (TDR) and gating capabilities. This enables engineers to identify and analyze impedance mismatches and faults over a defined time window, helping to troubleshoot and optimize RF systems effectively.

Another notable technology integrated into the HP 8753ET is its user-friendly interface that combines an easy-to-navigate display with direct access to measurement functions. This minimizes the learning curve for new users while enhancing productivity for experienced professionals.

The analyzer also supports various calibration techniques, ensuring accurate and repeatable measurements. The built-in calibration kits help improve the reliability of the results while compensating for any systematic errors associated with the test setup.

Additional characteristics of the HP 8753ET include its capability for both single-port and two-port measurements. This versatility allows users to conduct a complete array of S-parameter measurements with confidence.

Furthermore, the HP 8753ET features a storage function that allows users to save and recall measurement setups and results. This capability streamlines workflows and provides easy access to historical data for review or further analysis.

Ultimately, the HP 8753ET stands out as an essential tool for engineers in the RF and microwave domain. Its combination of high-performance measurement capabilities, an intuitive interface, and flexibility makes it a preferred choice for anyone seeking reliable and thorough network analysis in their testing and development projects.