Operating the Pump Motor in the STOPPED Display Mode | 5 |
|
|
TUBE ID |
Display shows |
TUBE ID = XXX |
ACTION |
Displays current |
TUBE ID |
1.Press the TUBE ID key. The display shows the last TUBE ID setting used, in inches or millimeters. Table
2.Use the UP and DOWN ARROW keys to enter a TUBE ID on the display. Once you start the pump motor, after selecting a mode (refer to Chapter 6), the pump will dispense fluid at a flow rate calculated based on the last TUBE ID displayed before the pump starts to run.
Using the UP ARROW key () will cause the TUBE ID value to scroll up through the available choices.
OR
Using the DOWN ARROW key () will cause the TUBE ID value to scroll down
through the available choices.
NOTE
If the red LED next to the CALIBRATE key is lit, the pump motor is running according to the last calibration complete (refer to Section 5.5).
If you press the TUBE ID key while the red LED is lit, the display shows “CALIBRATED TUBE!”.
Table 5-2. Tube Inner Diameters
| Tube Inner Diameters |
English | Metric |
1⁄32" | 1.0 mm |
|
|
1⁄16" | 2.0 mm |
|
|
1⁄8" | 3.0 mm |
|
|
3⁄16" | 4.0 mm |
|
|
1⁄4" | 5.0 mm |
|
|
5⁄16" | 6.0 mm |
|
|
| 7.0 mm |
|
|
| 8.0 mm |
|
|