Omega Speaker Systems FPU5-MT-220, FPU5-MT-110 manual Tips for Optimum Calibration Accuracy

Page 35

5Operating the Pump Motor in the STOPPED Display Mode

10.After you have finished dispensing, press the START/STOP key to stop the pump motor. Then measure the dispensed fluid’s volume using your calibrated beaker.

11.When you have determined the volume of the dispensed fluid, use the UP and DOWN ARROW keys to enter this volume on the display.

12.Once the display shows the correct volume of fluid dispensed, press the CALIBRATE key to complete the calibration. The display flashes and shows, “CALIBRATED!”, and the red LED next to the CALIBRATE key lights up.

13.Now you can select a mode (RPM, FLOW RATE, TIME or TOTAL VOLUME) (refer to Chapter 6). Once started, the fluid will be pumped according to the calibration.

NOTE

If you replace your tubing, make sure to press the TUBE ID key and select a different tube id size using the UP or DOWN ARROW keys.

NOTE

You can independently calibrate only one pump at a time. If you have two pumps attached to the motor (even if they are stacked), only one of the pumps can be calibrated.

For optimum FLOW RATE or TOTAL VOLUME accuracy, you will need to recalibrate the tubing as a result of:

Using a different tube size (inner diameter)

Using a different tubing material

Using a tube of a different age (calibration should be checked every 5-10 hours.)

Using a different back pressure in the tube

Changing the fluid being pumped

Changing the fluid’s temperature or density

Changing the RPM or FLOW RATE setting

Reversing the direction of pumping

Changing the clamp screw adjustment of the pump(s)

TIPS FOR OPTIMUM CALIBRATION ACCURACY:

Presented below are three useful techniques that will help you obtain optimum calibration accuracy.

Calibrate the unit at the speed you plan to use for normal pump operation.

When dispensing fluid during calibration, allow the pump to dispense the largest total volume that your measurement equipment supports.

When measuring the dispensed fluid during calibration, calculate the volume by measuring the fluid's mass with an accurate gram scale and dividing this mass by the fluid's density.

5-8

Image 35
Contents User’s Guide Servicing North America Unpacking Instructions Table of Peristaltic Pump Motor Peristaltic Pump Motor Chapter Specifications Spare Parts and Accessories 10-1Peristaltic Pump Motor RPM IndexIndex III Description FeaturesIntroduction Introduction Description Function Parts of the Pump MotorFront View Front ViewLeft Side View Left Side ViewRight Side View Right Side ViewMembrane Keypad Detail Membrane Keypad and DisplayFlow Rate key Membrane Keypad Details cont’dRear View Rear ViewUP and Down Arrow Keys Scrolling Limits Language Button Rear View cont’dParts of the Pump Motor Introduction Pump Motor Mounting Configurations Number of Pumps onGetting Started Attaching the PumpRefer to Figures 3-1 through Attaching a Single Pump Directly to the MotorPlacing the Second Pump on Top of the First Pump Stacking Two Pumps on One Side of the MotorLoading the Tubing Selecting TubingTubing Life See .2 to set the pumping direction Getting Started Introduction to the Two Display Modes Turning on the PowerUsing the ON/OFF Switch Stopped Display ModeStopped vs. Running Display Modes Running Display ModeDisplaying Pump Messages in Five Different Languages Stopped Display ModeOperating the Pump Motor Régler LA QTÉ English French Spanish German ItalianDefining the Direction of Pumping Left Pump Right Pump Selecting the Tubing Inner DiameterTube Inner Diameters English Metric Calibrating the Tubing Priming the TubingOperating the Pump Motor in the Stopped Display Mode Tips for Optimum Calibration Accuracy Preparing for Pumping in a Specific Mode Measuring the Temperature of a FluidTotal Volume Pumping in the RPM Mode Running Display ModeRPM Mode Shows the range of the RPM numbers you can work with RPM Mode Cont’d Pumping in the Flow Rate Mode Flow Rate ModeShows the ranges allowed for the various tubing sizes Flow Rate Ranges cont’d Display Xxx mL/min Flow Rate Mode Flow Rate Mode Pumping in the Time Mode Time ModeTime Range Display Xxxxxx Time Mode Time Mode Block Diagram Pumping in the Total Volume Mode Total Volume ModeTotal Volume Range Pumping in the Total Volume ModeOperating the Pump Motor in the Running Display Mode Total Volume Mode Cont’d Auxiliary Out Remote Monitor Preventing Tubing from Moving During Pump OperationOther Pumping Options Auxiliary In Remote Start/Stop 500Ω LEDSolution Troubleshooting GuideDisplay Error Messages Error Message ProblemRunning Display Mode and try to select the command again Error MessageLocked Rotor? Overcurrent Press the Total Volume key Short Circuit? Error Message Stopped Display Mode and refer to Section Other Pumping ProblemsProblem Solution Another pump, if necessary Troubleshooting Guide Theory of Operation Design SolutionMaintenance Technical DetailsTechnical Details Calibration Accuracy SpecificationsSpeed Average Flow RatesMotor Tubing MaterialsAverage Tubing Life ⁄32 300 ⁄16 250 100 ML Per Revolution Specifications Part Number Description Spare Parts and AccessoriesAccessories Spare PartsAccessories cont’d Peristatic Pump 10-4 WARRANTY/DISCLAIMER