6Operating the Pump Motor in the RUNNING Display Mode

6.Once the display shows the TOTAL VOLUME setting you desire, press the START/STOP key. As the fluid begins to dispense, the non-flashing display shows the calculated volume of dispensed fluid, as it occurs in time.

NOTE

In the TOTAL VOLUME Mode, the pump stops running after it dispenses the TOTAL VOLUME requested.

To stop pumping before that time, press the START/STOP key. The pump stops running, and the display returns to the STOPPED DISPLAY Mode.

When the pump stops, the display shows the calculated volume of fluid dispensed.

Do not be alarmed if this number is slightly larger than the TOTAL VOLUME you requested. Since the pump motor can not stop instantaneously, this number includes any small amount of fluid that escapes while the pump is in the process of stopping.

Table 6-4 shows how other keys and buttons can be used while pumping in the TOTAL VOLUME mode.

6-18

Image 55
Contents User’s Guide Servicing North America Unpacking Instructions Table of Peristaltic Pump Motor Peristaltic Pump Motor Chapter Specifications Spare Parts and Accessories 10-1Peristaltic Pump Motor RPM IndexIndex III Introduction FeaturesDescription Introduction Description Function Parts of the Pump MotorFront View Front ViewLeft Side View Left Side ViewRight Side View Right Side ViewMembrane Keypad Detail Membrane Keypad and DisplayFlow Rate key Membrane Keypad Details cont’dUP and Down Arrow Keys Scrolling Limits Rear ViewRear View Language Button Rear View cont’dParts of the Pump Motor Introduction Pump Motor Mounting Configurations Number of Pumps onGetting Started Attaching the PumpRefer to Figures 3-1 through Attaching a Single Pump Directly to the MotorPlacing the Second Pump on Top of the First Pump Stacking Two Pumps on One Side of the MotorTubing Life Selecting TubingLoading the Tubing See .2 to set the pumping direction Getting Started Introduction to the Two Display Modes Turning on the PowerUsing the ON/OFF Switch Stopped Display ModeStopped vs. Running Display Modes Running Display ModeOperating the Pump Motor Stopped Display ModeDisplaying Pump Messages in Five Different Languages Régler LA QTÉ English French Spanish German ItalianDefining the Direction of Pumping Left Pump Right Pump Selecting the Tubing Inner DiameterTube Inner Diameters English Metric Calibrating the Tubing Priming the TubingOperating the Pump Motor in the Stopped Display Mode Tips for Optimum Calibration Accuracy Preparing for Pumping in a Specific Mode Measuring the Temperature of a FluidTotal Volume RPM Mode Running Display ModePumping in the RPM Mode Shows the range of the RPM numbers you can work with RPM Mode Cont’d Pumping in the Flow Rate Mode Flow Rate ModeShows the ranges allowed for the various tubing sizes Flow Rate Ranges cont’d Display Xxx mL/min Flow Rate Mode Flow Rate Mode Pumping in the Time Mode Time ModeTime Range Display Xxxxxx Time Mode Time Mode Block Diagram Pumping in the Total Volume Mode Total Volume ModeTotal Volume Range Pumping in the Total Volume ModeOperating the Pump Motor in the Running Display Mode Total Volume Mode Cont’d Other Pumping Options Preventing Tubing from Moving During Pump OperationAuxiliary Out Remote Monitor Auxiliary In Remote Start/Stop 500Ω LEDSolution Troubleshooting GuideDisplay Error Messages Error Message ProblemRunning Display Mode and try to select the command again Error MessageLocked Rotor? Overcurrent Press the Total Volume key Short Circuit? Error Message Stopped Display Mode and refer to Section Other Pumping ProblemsProblem Solution Another pump, if necessary Troubleshooting Guide Theory of Operation Design SolutionMaintenance Technical DetailsTechnical Details Calibration Accuracy SpecificationsSpeed Average Flow RatesMotor Tubing MaterialsAverage Tubing Life ⁄32 300 ⁄16 250 100 ML Per Revolution Specifications Part Number Description Spare Parts and AccessoriesAccessories Spare PartsAccessories cont’d Peristatic Pump 10-4 WARRANTY/DISCLAIMER