Omega Speaker Systems FPU5-MT-220, FPU5-MT-110 manual Error Message

Page 61

7Troubleshooting Guide

Error Message

Problem

Illegal Op Code (continued)

B

There may be significant electrical noise affecting the pump motor.

Solution

3.If the line voltage is inappropriate for the pump motor, move it to another line

voltage source.

B

Check the pumping environment for devices that could be creating electrical noise. If possible, shut off these devices or move the pump to another location.

C

There may be a problem with the microprocessor.

C

Turn the unit’s power off and on, adjusting the settings, and attempt to resume pumping.

Is pump running?

You have selected a command

 

that requires the pump motor

 

to be running.

1.Press the START/STOP key to start the motor.

2.Verify that the display is in the

RUNNING Display Mode and try to select the command again.

Locked Rotor?

The motor should be running, but

 

 

 

no pulse interrupts are occurring,

 

 

 

so it is not running.

 

 

 

To recover from a locked rotor

 

 

 

condition, first turn the power switch

 

 

 

off and then back on. Then attempt

 

 

 

a solution below.

 

 

 

A

A

 

 

The motor shaft may be locked.

1.

The tubing may be causing a

 

 

 

jam. Observe how the tubing

 

 

 

is loaded in the pump head

 

 

 

and, if necessary, reload it.

 

 

2.

Detach the pump(s) and check

 

 

 

for a mechanical jam.

 

 

 

Substitute another pump, if

 

 

 

necessary.

 

 

3.

Turn the unit’s power off and

 

 

 

on, adjust the settings, and

 

 

 

attempt to resume pumping.

7-2

Image 61
Contents User’s Guide Servicing North America Unpacking Instructions Table of Peristaltic Pump Motor Peristaltic Pump Motor Chapter Specifications Spare Parts and Accessories 10-1Peristaltic Pump Motor RPM IndexIndex III Introduction FeaturesDescription Introduction Front View Parts of the Pump MotorFront View Description FunctionLeft Side View Left Side ViewRight Side View Right Side ViewMembrane Keypad Detail Membrane Keypad and DisplayFlow Rate key Membrane Keypad Details cont’dUP and Down Arrow Keys Scrolling Limits Rear ViewRear View Language Button Rear View cont’dParts of the Pump Motor Getting Started Pump Motor Mounting Configurations Number of Pumps onAttaching the Pump IntroductionRefer to Figures 3-1 through Attaching a Single Pump Directly to the MotorPlacing the Second Pump on Top of the First Pump Stacking Two Pumps on One Side of the MotorTubing Life Selecting TubingLoading the Tubing See .2 to set the pumping direction Getting Started Using the ON/OFF Switch Turning on the PowerStopped Display Mode Introduction to the Two Display ModesStopped vs. Running Display Modes Running Display ModeOperating the Pump Motor Stopped Display ModeDisplaying Pump Messages in Five Different Languages Régler LA QTÉ English French Spanish German ItalianDefining the Direction of Pumping Left Pump Right Pump Selecting the Tubing Inner DiameterTube Inner Diameters English Metric Calibrating the Tubing Priming the TubingOperating the Pump Motor in the Stopped Display Mode Tips for Optimum Calibration Accuracy Preparing for Pumping in a Specific Mode Measuring the Temperature of a FluidTotal Volume RPM Mode Running Display ModePumping in the RPM Mode Shows the range of the RPM numbers you can work with RPM Mode Cont’d Pumping in the Flow Rate Mode Flow Rate ModeShows the ranges allowed for the various tubing sizes Flow Rate Ranges cont’d Display Xxx mL/min Flow Rate Mode Flow Rate Mode Pumping in the Time Mode Time ModeTime Range Display Xxxxxx Time Mode Time Mode Block Diagram Pumping in the Total Volume Mode Total Volume ModeTotal Volume Range Pumping in the Total Volume ModeOperating the Pump Motor in the Running Display Mode Total Volume Mode Cont’d Other Pumping Options Preventing Tubing from Moving During Pump OperationAuxiliary Out Remote Monitor Auxiliary In Remote Start/Stop 500Ω LEDDisplay Error Messages Troubleshooting GuideError Message Problem SolutionRunning Display Mode and try to select the command again Error MessageLocked Rotor? Overcurrent Press the Total Volume key Short Circuit? Error Message Stopped Display Mode and refer to Section Other Pumping ProblemsProblem Solution Another pump, if necessary Troubleshooting Guide Maintenance Design SolutionTechnical Details Theory of OperationTechnical Details Calibration Accuracy SpecificationsMotor Average Flow RatesTubing Materials SpeedAverage Tubing Life ⁄32 300 ⁄16 250 100 ML Per Revolution Specifications Accessories Spare Parts and AccessoriesSpare Parts Part Number DescriptionAccessories cont’d Peristatic Pump 10-4 WARRANTY/DISCLAIMER