Teledyne 450H Control Inputs, Output, Description, Normal Operating State, 1 DIGITAL OUTPUTS

Page 20
Output

 

Output

 

 

Description

 

 

Normal Operating State

 

 

 

 

 

 

 

1

 

Power OK

ON

2

 

 

Diagnostics Mode(Instrument not

 

OFF

 

 

 

 

monitoring)

 

 

 

3

 

 

Temperature Fault

OFF

4

 

 

Pressure Fault

OFF

5

 

 

UV Lamp Fault

OFF

6

 

 

Flow Fault

OFF

7

 

 

Performing Auto-Zero Calibration

OFF

8-11

 

Not Used

N/A

12

 

 

System OK(no faults)

ON

TABLE 1.1 DIGITAL OUTPUTS

Each digital output is configured as a Collector/Emitter pair. The labels are ‘C’ and ‘E’ for these contacts, respectively. Figure 1.2 below shows the most common way of connecting the digital outputs to an external device such as PLC. Note: Most devices, such as PLC’s, have internal provision for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, external dropping resistors must be used to limit the current through the transistor output to 50mA or less.

FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS

1.6.4 Control Inputs

There are 6 digital control inputs on the rear panel; they are labeled ‘CONTROL IN’ (See Figure 1.4.) The control inputs are used for remote control of the M450H by a device such as a PLC. These inputs are triggered by providing a contact closure or low impedance current path between the + and ground contacts. This can be done by using a mechanical switch or isolated transistor type output from another device, such as a PLC. Never connect a voltage level output from another device to these contacts. The functions of the control inputs is summarized below in Table 1.2:

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 20

Image 20
Contents Printed 05/15/06 MODEL 450H OZONE MONITOREMAIL api-sales@teledyne.com SAFETY MESSAGES TABLE OF CONTENTS 3.0 SETUP MODE4.0 MAINTENANCE 5.0 ADJUSTMENTS5.4 CURRENT LOOP CALIBRATION APPENDIX A - ELECTRICAL SCHEMATIC INDEX5.3 A/D - D/A CALIBRATION PROCEDURE… FIGURES TABLES TABLE 1.1 DIGITAL OUTPUTSTABLE 1.2 CONTROL INPUTS TABLE 1.3 FINAL TEST AND CALIBRATION VALUES1.0INTRODUCTION 1.1Preface1.2 WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION02024c COVERAGE1.3 Principle of Operation Page 1.4 Specifications 1.5 Installation 19” Rack Mount EnclosurePage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS 1.6Electrical I/O Connections1.6.1Analog Output 1.6.2 Relay OutputsFIGURE 1.5 - CONNECTING DIGITAL OUTPUTS 1.6.4 Control InputsOutput Description1.6.5 RS232 Output InputCONTROL INPUTS Zero CalibrateFIGURE 1.6 - NEMA4X ASSEMBLY LAYOUT FIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAMU V L A M P M O T H E R B V / F C P U A C L I N E D C P O W E R SOFigure 1.8 - 19” Rack Assembly Layout POWER SUPPLY MODULE1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.5 Flow Switch 1.8 Options1.8.1 Auto-ZeroOption 1.8.2 Isolated Current Output 4-20mAor 2-20mA2.1Key features 2.1.2 Concentration Alarms2.1.4 E2 ROM backup of software configuration 2.0OPERATION2.2 Front Panel Display 2.2.1 Front panel display fieldsFIGURE 2.1 - MODEL 450H FRONT PANEL Mode FieldMessage/Test Measurement Field Test MessageTEST MEASUREMENTS Warning Message2.3 Software Operation 2.2.2 Status LED’s2.3.1 Main Menu State2.3.3 Alarm Status Menu FIGURE 2.3 - ALARM STATUS MENUClearing Alarms Audible BeeperPage 3.0 SETUP MODE FIGURE 3.1 - SETUP MENU3.1 Setting the Concentration Alarms ALRM 3.2 Setting the Concentration Units UNIT3.3 Setting the time-of-dayand date CLK HI ALARM OFF3.4 Setting the RS-232baud rate COMM 3.5 Setup variables VARS 0 LATCH_ALARMS1 ALARM BEEPER 3 AZERO ENABLE8 GAS MOLECULAR WEIGHT 10 TPC ENABLE7 AZERO HOLD OFF 9 CONCENTRATION RANGE3.5Diagnostics DIAG 3.5.1Zero Calibration3.5.2 D/A Calibration 3.5.3 Analog Output4.0Maintenance 4.2Cleaning Exterior Surfaces of the M450H4.3 Cleaning the Optical Cell Filter Replacement ProcedureFIGURE 4.1 MEASUREMENT CELL DISASSEMBLY 4.4 Degree of Protection5.0 ADJUSTMENTS Switch Settings5.1 Calibration 5.2Changing the analog output rangeFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.3 A/D - D/A Calibration Procedure1.Press SETUP-DIAG 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page APPENDIX B - SPARE PARTS LIST DESCRIPTIONPage Title APPENDIX C - ELECTRICAL SCHEMATIC INDEXDrawing Number