Teledyne 450H instruction manual Operation verification

Page 24
1.7 Operation verification

1.7 Operation verification

The Model 450H Monitor is now ready for operation.

1.Read Sections 1.3 and all of Section 2 of the manual to understand the Monitor's operation.

2.Turn on the power by pressing the on/off switch (see Figure 2.1). The display should turn on and green (sample) status LED should be energized. The yellow "fault" light may also be on until the temperatures and voltages are within operating limits. Clear the fault messages.

3.Sample Flow - Adjust to 0.5-2.0 L/min using needle valve (See Figure 1.6)

4.After a 20 minute warm-up, review the TEST function values in the front panel display by pushing the keyboard button labeled TST>. Not every TEST function is a diagnostic of correct monitor operation, therefore TEST functions not covered below can be ignored for now.

5.O3 REF, O3 MEAS - TEST function values should be between 4200 mV and 4700 mV.

6.Pressure – 14.5-15.2 psia at sea level. Other values will be displayed depending on altitude of monitor.

7.Sample Temp - 20 -50oC

8.Box Temp - Ambient +10 oC

9.If the TEST functions are within the limits given above the instrument should function correctly. If there is a problem please read the manual and check your set-up.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 24

Image 24
Contents MODEL 450H OZONE MONITOR EMAIL api-sales@teledyne.comPrinted 05/15/06 SAFETY MESSAGES TABLE OF CONTENTS 3.0 SETUP MODE4.0 MAINTENANCE 5.0 ADJUSTMENTSAPPENDIX A - ELECTRICAL SCHEMATIC INDEX 5.3 A/D - D/A CALIBRATION PROCEDURE…5.4 CURRENT LOOP CALIBRATION FIGURES TABLES TABLE 1.1 DIGITAL OUTPUTSTABLE 1.2 CONTROL INPUTS TABLE 1.3 FINAL TEST AND CALIBRATION VALUES1.0INTRODUCTION 1.1Preface1.2 WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION02024c COVERAGE1.3 Principle of Operation Page 1.4 Specifications 1.5 Installation 19” Rack Mount EnclosurePage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS 1.6Electrical I/O Connections1.6.1Analog Output 1.6.2 Relay OutputsFIGURE 1.5 - CONNECTING DIGITAL OUTPUTS 1.6.4 Control InputsOutput Description1.6.5 RS232 Output InputCONTROL INPUTS Zero CalibrateFIGURE 1.6 - NEMA4X ASSEMBLY LAYOUT FIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAMU V L A M P M O T H E R B V / F C P U A C L I N E D C P O W E R SOFigure 1.8 - 19” Rack Assembly Layout POWER SUPPLY MODULE1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.5 Flow Switch 1.8 Options1.8.1 Auto-ZeroOption 1.8.2 Isolated Current Output 4-20mAor 2-20mA2.1Key features 2.1.2 Concentration Alarms2.1.4 E2 ROM backup of software configuration 2.0OPERATION2.2 Front Panel Display 2.2.1 Front panel display fieldsFIGURE 2.1 - MODEL 450H FRONT PANEL Mode FieldMessage/Test Measurement Field Test MessageTEST MEASUREMENTS Warning Message2.3 Software Operation 2.2.2 Status LED’s2.3.1 Main Menu State2.3.3 Alarm Status Menu FIGURE 2.3 - ALARM STATUS MENUClearing Alarms Audible BeeperPage 3.0 SETUP MODE FIGURE 3.1 - SETUP MENU3.1 Setting the Concentration Alarms ALRM 3.2 Setting the Concentration Units UNIT3.3 Setting the time-of-dayand date CLK HI ALARM OFF3.4 Setting the RS-232baud rate COMM 3.5 Setup variables VARS 0 LATCH_ALARMS1 ALARM BEEPER 3 AZERO ENABLE8 GAS MOLECULAR WEIGHT 10 TPC ENABLE7 AZERO HOLD OFF 9 CONCENTRATION RANGE3.5Diagnostics DIAG 3.5.1Zero Calibration3.5.2 D/A Calibration 3.5.3 Analog Output4.0Maintenance 4.2Cleaning Exterior Surfaces of the M450H4.3 Cleaning the Optical Cell Filter Replacement ProcedureFIGURE 4.1 MEASUREMENT CELL DISASSEMBLY 4.4 Degree of Protection5.0 ADJUSTMENTS Switch Settings5.1 Calibration 5.2Changing the analog output range5.3 A/D - D/A Calibration Procedure 1.Press SETUP-DIAGFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page APPENDIX B - SPARE PARTS LIST DESCRIPTIONPage APPENDIX C - ELECTRICAL SCHEMATIC INDEX Drawing NumberTitle