Teledyne 450H Alarm Status Menu, 3 - ALARM STATUS MENU, Clearing Alarms, Audible Beeper

Page 31
2.3.3 Alarm Status Menu

When the MSG button is displayed, it indicates that one or more warnings have been issued. To view the warning messages, press the MSG button. Repeatedly pressing this button will cycle through all available warning messages. Pressing CLR will clear the warning message from the display. Note: If the conditions that caused the warning messages are still in effect, the warning messages will re-appear after they have been cleared.

The ALRM button is displayed when concentration alarms have been triggered. See Section 2.3.3 for operation of the Alarm menu.

2.3.3 Alarm Status Menu

Pressing the ALRM button from the main menu when alarms have been triggered will bring up the alarm status menu as shown below in Figure 2.3.

 

ALARM STATUS

 

HI=ACTIVE, HI HI=OK

 

 

 

 

 

 

 

 

 

 

 

CLR

 

 

 

BEEP

 

EXIT

 

 

 

 

 

 

 

 

 

FIGURE 2.3 - ALARM STATUS MENU

Clearing Alarms

If the monitor is configured for Latching type alarms, and either of the alarms can be cleared (ie. The concentration has dropped below the alarm threshold,) then a CLR button will appear as shown in Figure 2.3 above. The CLR button will clear the alarm condition.

If the monitor is configured for Non-Latching alarms, the CLR button will not appear, and all alarms will automatically clear when the concentration drops below the alarm threshold.

Audible Beeper

An audible beeper is sounded when any alarm is activated. A slow beep indicates that the ‘HI’ alarm has been triggered, and a fast beep indicates that the ‘HI-HI’ alarm has been triggered. Pressing the BEEP button from the alarm menu will silence the audible beeper for 5 minutes. Note: Pressing the BEEP button does not clear any of the alarms, it simply silences the audible alarm. If the alarms are not cleared, the audible alarm will automatically resume in 5 minutes.

See Sections 3.1 and 3.5 for details on configuring concentration alarms.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 31

Image 31
Contents EMAIL api-sales@teledyne.com MODEL 450H OZONE MONITORPrinted 05/15/06 SAFETY MESSAGES 5.0 ADJUSTMENTS TABLE OF CONTENTS3.0 SETUP MODE 4.0 MAINTENANCE5.3 A/D - D/A CALIBRATION PROCEDURE… APPENDIX A - ELECTRICAL SCHEMATIC INDEX5.4 CURRENT LOOP CALIBRATION FIGURES TABLE 1.3 FINAL TEST AND CALIBRATION VALUES TABLESTABLE 1.1 DIGITAL OUTPUTS TABLE 1.2 CONTROL INPUTS1.1Preface 1.0INTRODUCTIONCOVERAGE 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c1.3 Principle of Operation Page 1.5 Installation 19” Rack Mount Enclosure 1.4 SpecificationsPage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS 1.6.2 Relay Outputs FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS1.6Electrical I/O Connections 1.6.1Analog OutputDescription FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS1.6.4 Control Inputs OutputZero Calibrate 1.6.5 RS232 OutputInput CONTROL INPUTSD C P O W E R SO FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUTFIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM U V L A M P M O T H E R B V / F C P U A C L I N EPOWER SUPPLY MODULE Figure 1.8 - 19” Rack Assembly Layout1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.2 Isolated Current Output 4-20mAor 2-20mA 1.8.5 Flow Switch1.8 Options 1.8.1 Auto-ZeroOption2.0OPERATION 2.1Key features2.1.2 Concentration Alarms 2.1.4 E2 ROM backup of software configurationMode Field 2.2 Front Panel Display2.2.1 Front panel display fields FIGURE 2.1 - MODEL 450H FRONT PANELWarning Message Message/Test Measurement FieldTest Message TEST MEASUREMENTSState 2.3 Software Operation2.2.2 Status LED’s 2.3.1 Main MenuAudible Beeper 2.3.3 Alarm Status MenuFIGURE 2.3 - ALARM STATUS MENU Clearing AlarmsPage FIGURE 3.1 - SETUP MENU 3.0 SETUP MODEHI ALARM OFF 3.1 Setting the Concentration Alarms ALRM3.2 Setting the Concentration Units UNIT 3.3 Setting the time-of-dayand date CLK3.4 Setting the RS-232baud rate COMM 3 AZERO ENABLE 3.5 Setup variables VARS0 LATCH_ALARMS 1 ALARM BEEPER9 CONCENTRATION RANGE 8 GAS MOLECULAR WEIGHT10 TPC ENABLE 7 AZERO HOLD OFF3.5.3 Analog Output 3.5Diagnostics DIAG3.5.1Zero Calibration 3.5.2 D/A CalibrationFilter Replacement Procedure 4.0Maintenance4.2Cleaning Exterior Surfaces of the M450H 4.3 Cleaning the Optical Cell4.4 Degree of Protection FIGURE 4.1 MEASUREMENT CELL DISASSEMBLY5.2Changing the analog output range 5.0 ADJUSTMENTSSwitch Settings 5.1 Calibration1.Press SETUP-DIAG 5.3 A/D - D/A Calibration ProcedureFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page DESCRIPTION APPENDIX B - SPARE PARTS LISTPage Drawing Number APPENDIX C - ELECTRICAL SCHEMATIC INDEXTitle