Teledyne 450H instruction manual Current Loop Calibration

Page 43
5.4 Current Loop Calibration

5.4 Current Loop Calibration

A current loop option can be ordered that will provide a 0-20mA or 4-20mA output on the analog output. This calibration must be repeated every time an A/D - D/A calibration is performed. To calibrate the current output, perform the following steps:

1.Perform an A/D - D/A calibration as outlined in Section 5.3.

2.Connect a Multimeter capable of measuring milliamperes to the analog output on the rear panel. Note: When measuring the current output with a multimeter or similar low-impedance current measuring device, a 400- 450 ohm resistor must be placed in series with the meter to simulate a load. Failure to do this will result in erroneous readings.

3.From the front panel, press SETUP-DIAG. Press NEXT until D/A CALIBRATION appears and press ENTR. Press CFG and the properties for analog output Channel 0 will be displayed on the top line. The display should show something like:

DIAG D/A

 

 

 

0) O3_CONC_1,CURR, NOT CAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NEXT

 

CAL

 

SET

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

4.This indicates that channel 0 is setup for current output and has not been calibrated. If the display shows VOLT instead of CURR then the channel must be setup for current output. To do this, press SET, select CURR as the output type and press ENTR.

5.Next press CAL to begin the calibration. At this point the display should show:

DIAG D/A

 

 

 

CAL O3_CONC_1,CURR, ZERO

 

 

 

 

 

 

 

 

 

 

 

U100

 

UP10

 

 

UP

 

DOWN DN10 D100 ENTR EXIT

 

 

 

 

 

 

 

 

 

 

6.The zero point for current output can now be set. Pressing UP, UP10 and U100 will step the zero point up in increments of 1, 10 and 100 steps. The

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 43

Image 43
Contents EMAIL api-sales@teledyne.com MODEL 450H OZONE MONITORPrinted 05/15/06 SAFETY MESSAGES 5.0 ADJUSTMENTS TABLE OF CONTENTS3.0 SETUP MODE 4.0 MAINTENANCE5.3 A/D - D/A CALIBRATION PROCEDURE… APPENDIX A - ELECTRICAL SCHEMATIC INDEX5.4 CURRENT LOOP CALIBRATION FIGURES TABLE 1.3 FINAL TEST AND CALIBRATION VALUES TABLESTABLE 1.1 DIGITAL OUTPUTS TABLE 1.2 CONTROL INPUTS1.1Preface 1.0INTRODUCTIONCOVERAGE 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c1.3 Principle of Operation Page 1.5 Installation 19” Rack Mount Enclosure 1.4 SpecificationsPage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS 1.6.2 Relay Outputs FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS1.6Electrical I/O Connections 1.6.1Analog OutputDescription FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS1.6.4 Control Inputs OutputZero Calibrate 1.6.5 RS232 OutputInput CONTROL INPUTSD C P O W E R SO FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUTFIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM U V L A M P M O T H E R B V / F C P U A C L I N EPOWER SUPPLY MODULE Figure 1.8 - 19” Rack Assembly Layout1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.2 Isolated Current Output 4-20mAor 2-20mA 1.8.5 Flow Switch1.8 Options 1.8.1 Auto-ZeroOption2.0OPERATION 2.1Key features2.1.2 Concentration Alarms 2.1.4 E2 ROM backup of software configurationMode Field 2.2 Front Panel Display2.2.1 Front panel display fields FIGURE 2.1 - MODEL 450H FRONT PANELWarning Message Message/Test Measurement FieldTest Message TEST MEASUREMENTSState 2.3 Software Operation2.2.2 Status LED’s 2.3.1 Main MenuAudible Beeper 2.3.3 Alarm Status MenuFIGURE 2.3 - ALARM STATUS MENU Clearing AlarmsPage FIGURE 3.1 - SETUP MENU 3.0 SETUP MODEHI ALARM OFF 3.1 Setting the Concentration Alarms ALRM3.2 Setting the Concentration Units UNIT 3.3 Setting the time-of-dayand date CLK3.4 Setting the RS-232baud rate COMM 3 AZERO ENABLE 3.5 Setup variables VARS0 LATCH_ALARMS 1 ALARM BEEPER9 CONCENTRATION RANGE 8 GAS MOLECULAR WEIGHT10 TPC ENABLE 7 AZERO HOLD OFF3.5.3 Analog Output 3.5Diagnostics DIAG3.5.1Zero Calibration 3.5.2 D/A CalibrationFilter Replacement Procedure 4.0Maintenance4.2Cleaning Exterior Surfaces of the M450H 4.3 Cleaning the Optical Cell4.4 Degree of Protection FIGURE 4.1 MEASUREMENT CELL DISASSEMBLY5.2Changing the analog output range 5.0 ADJUSTMENTSSwitch Settings 5.1 Calibration1.Press SETUP-DIAG 5.3 A/D - D/A Calibration ProcedureFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page DESCRIPTION APPENDIX B - SPARE PARTS LISTPage Drawing Number APPENDIX C - ELECTRICAL SCHEMATIC INDEXTitle