Degree of Protection, 1 MEASUREMENT CELL DISASSEMBLY
Specs
Parts list
Concentration Alarms
Warranty
Maintenance
3.5Diagnostics DIAG
Setup Mode
Adjustments
Weight
Safety
Page 40
Page 39
Page 41
Image 40
Page 39
Page 41
Contents
EMAIL api-sales@teledyne.com
MODEL 450H OZONE MONITOR
Printed 05/15/06
SAFETY MESSAGES
TABLE OF CONTENTS
3.0 SETUP MODE
4.0 MAINTENANCE
5.0 ADJUSTMENTS
5.3 A/D - D/A CALIBRATION PROCEDURE…
APPENDIX A - ELECTRICAL SCHEMATIC INDEX
5.4 CURRENT LOOP CALIBRATION
FIGURES
TABLES
TABLE 1.1 DIGITAL OUTPUTS
TABLE 1.2 CONTROL INPUTS
TABLE 1.3 FINAL TEST AND CALIBRATION VALUES
1.0INTRODUCTION
1.1Preface
1.2 WARRANTY POLICY
ADVANCED POLLUTION INSTRUMENTATION DIVISION
02024c
COVERAGE
1.3 Principle of Operation
Page
1.4 Specifications
1.5 Installation 19” Rack Mount Enclosure
Page
5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line
1.6 Installation NEMA 4X Wall Mount Enclosure
ozone
FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS
FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS
FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS
FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS
1.6Electrical I/O Connections
1.6.1Analog Output
1.6.2 Relay Outputs
FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS
1.6.4 Control Inputs
Output
Description
1.6.5 RS232 Output
Input
CONTROL INPUTS
Zero Calibrate
FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUT
FIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM
U V L A M P M O T H E R B V / F C P U A C L I N E
D C P O W E R SO
Figure 1.8 - 19” Rack Assembly Layout
POWER SUPPLY MODULE
1.7 Operation verification
TABLE 1.3 FINAL TEST AND CALIBRATION VALUES
1.8.5 Flow Switch
1.8 Options
1.8.1 Auto-ZeroOption
1.8.2 Isolated Current Output 4-20mAor 2-20mA
2.1Key features
2.1.2 Concentration Alarms
2.1.4 E2 ROM backup of software configuration
2.0OPERATION
2.2 Front Panel Display
2.2.1 Front panel display fields
FIGURE 2.1 - MODEL 450H FRONT PANEL
Mode Field
Message/Test Measurement Field
Test Message
TEST MEASUREMENTS
Warning Message
2.3 Software Operation
2.2.2 Status LED’s
2.3.1 Main Menu
State
2.3.3 Alarm Status Menu
FIGURE 2.3 - ALARM STATUS MENU
Clearing Alarms
Audible Beeper
Page
3.0 SETUP MODE
FIGURE 3.1 - SETUP MENU
3.1 Setting the Concentration Alarms ALRM
3.2 Setting the Concentration Units UNIT
3.3 Setting the time-of-dayand date CLK
HI ALARM OFF
3.4 Setting the RS-232baud rate COMM
3.5 Setup variables VARS
0 LATCH_ALARMS
1 ALARM BEEPER
3 AZERO ENABLE
8 GAS MOLECULAR WEIGHT
10 TPC ENABLE
7 AZERO HOLD OFF
9 CONCENTRATION RANGE
3.5Diagnostics DIAG
3.5.1Zero Calibration
3.5.2 D/A Calibration
3.5.3 Analog Output
4.0Maintenance
4.2Cleaning Exterior Surfaces of the M450H
4.3 Cleaning the Optical Cell
Filter Replacement Procedure
FIGURE 4.1 MEASUREMENT CELL DISASSEMBLY
4.4 Degree of Protection
5.0 ADJUSTMENTS
Switch Settings
5.1 Calibration
5.2Changing the analog output range
1.Press SETUP-DIAG
5.3 A/D - D/A Calibration Procedure
FIGURE 4.2- ANALOG OUTPUT JUMPERS
5.4 Current Loop Calibration
zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter
APPENDIX A - SOFTWARE MENU TREE
Page
APPENDIX B - SPARE PARTS LIST
DESCRIPTION
Page
Drawing Number
APPENDIX C - ELECTRICAL SCHEMATIC INDEX
Title
Related pages
All Assembly page
Ribbon Assembly for SATO CL6XXe
Swing Spout Assembly for American Standard 7191 Series
VAC Reversing Electric Relay Panel Assembly for American Dryer Corp AD-25V
Assembly for Weber 1500
Step Five Setting The Chain Latch Assembly for Wayne 3514
Cassette Cover Assembly for Panasonic AJ-SDC615E
Assembly\Disassembly for Kodak 700
Assembly for Fisher-Price T3857
Assembly for Grizzly T10131
Mounting the DRS-10 Switch Assembly to a Wall for Shure DRS-10
What tools do I need for installing the canopy? Find out in our
Hunter Fan canopy section
.
Top
Page
Image
Contents