Teledyne 450H instruction manual Principle of Operation

Page 9
1.3 Principle of Operation

All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.

1.3 Principle of Operation

The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O3 molecule. The Model 450H uses a mercury lamp constructed so that a large majority of the light emitted is at the 254nm wavelength. Light from the lamp shines through an absorption cell through wich the sample gas being measured is passed. The ratio of the intensity of light passing through the gas to a reference measurement which does not pass through the gas forms the ratio I/Io. This ratio forms the basis for the calculation of the ozone concentration.

The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.

 

 

106

 

Τ

14.695psi

 

Ι

CO3 = −

 

×

 

×

 

× ln

 

 

α × l

273o Κ

Ρ

Ιo

Where:

 

 

 

 

 

 

 

 

 

 

I

=

Intensity of light passed through the sample

Io

=

Intensity of light through sample free of ozone

α= absorption coefficient l = path length

CO3 = concentration of ozone in parts per million

T= sample temperature in degrees Kelvin

P= pressure in pounds per square inch (absolute)

As can be seen the concentration of ozone depends on more than the intensity ratio. Temperature and pressure influence the density of the sample. The density changes the number of ozone molecules in the absorption cell which impacts the amount of light removed from the light beam. These effects are addressed by directly measuring temperature and pressure and including their actual values in the calculation. The absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254 nm light. Most current measurements place this value at 308 cm-1atm-1at STP. The value of this number reflects the fact that ozone is a very efficient absorber of UV radiation which is why stratospheric ozone protects the life forms lower in the atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path length determines how many molecules are present in the column of gas in the absorption cell.

The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution. The digitized signal, along with the other variables, are used by the CPU to compute the concentration using the above formula.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 9

Image 9
Contents MODEL 450H OZONE MONITOR EMAIL api-sales@teledyne.comPrinted 05/15/06 SAFETY MESSAGES 3.0 SETUP MODE TABLE OF CONTENTS4.0 MAINTENANCE 5.0 ADJUSTMENTSAPPENDIX A - ELECTRICAL SCHEMATIC INDEX 5.3 A/D - D/A CALIBRATION PROCEDURE…5.4 CURRENT LOOP CALIBRATION FIGURES TABLE 1.1 DIGITAL OUTPUTS TABLES TABLE 1.2 CONTROL INPUTS TABLE 1.3 FINAL TEST AND CALIBRATION VALUES1.1Preface 1.0INTRODUCTIONADVANCED POLLUTION INSTRUMENTATION DIVISION 1.2 WARRANTY POLICY02024c COVERAGE1.3 Principle of Operation Page 1.5 Installation 19” Rack Mount Enclosure 1.4 SpecificationsPage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS 1.6Electrical I/O Connections FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS1.6.1Analog Output 1.6.2 Relay Outputs1.6.4 Control Inputs FIGURE 1.5 - CONNECTING DIGITAL OUTPUTSOutput DescriptionInput 1.6.5 RS232 OutputCONTROL INPUTS Zero CalibrateFIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUTU V L A M P M O T H E R B V / F C P U A C L I N E D C P O W E R SOPOWER SUPPLY MODULE Figure 1.8 - 19” Rack Assembly Layout1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8 Options 1.8.5 Flow Switch1.8.1 Auto-ZeroOption 1.8.2 Isolated Current Output 4-20mAor 2-20mA2.1.2 Concentration Alarms 2.1Key features2.1.4 E2 ROM backup of software configuration 2.0OPERATION2.2.1 Front panel display fields 2.2 Front Panel DisplayFIGURE 2.1 - MODEL 450H FRONT PANEL Mode FieldTest Message Message/Test Measurement FieldTEST MEASUREMENTS Warning Message2.2.2 Status LED’s 2.3 Software Operation2.3.1 Main Menu StateFIGURE 2.3 - ALARM STATUS MENU 2.3.3 Alarm Status MenuClearing Alarms Audible BeeperPage FIGURE 3.1 - SETUP MENU 3.0 SETUP MODE3.2 Setting the Concentration Units UNIT 3.1 Setting the Concentration Alarms ALRM3.3 Setting the time-of-dayand date CLK HI ALARM OFF3.4 Setting the RS-232baud rate COMM 0 LATCH_ALARMS 3.5 Setup variables VARS1 ALARM BEEPER 3 AZERO ENABLE10 TPC ENABLE 8 GAS MOLECULAR WEIGHT7 AZERO HOLD OFF 9 CONCENTRATION RANGE3.5.1Zero Calibration 3.5Diagnostics DIAG3.5.2 D/A Calibration 3.5.3 Analog Output4.2Cleaning Exterior Surfaces of the M450H 4.0Maintenance4.3 Cleaning the Optical Cell Filter Replacement Procedure4.4 Degree of Protection FIGURE 4.1 MEASUREMENT CELL DISASSEMBLYSwitch Settings 5.0 ADJUSTMENTS5.1 Calibration 5.2Changing the analog output range5.3 A/D - D/A Calibration Procedure 1.Press SETUP-DIAGFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page DESCRIPTION APPENDIX B - SPARE PARTS LISTPage APPENDIX C - ELECTRICAL SCHEMATIC INDEX Drawing NumberTitle