This variable determines length of time that zero gas is admitted into the measurement cell before the zero calibration is performed. This value must be set to a long enough interval to ensure that the cell is purged of all ozone before the calibration is performed. The defaul value is 20 seconds.
7) AZERO_HOLD_OFF
This variable determines the length of time the instrument waits after the
8) GAS_MOLECULAR_WEIGHT
This variable is the molecular weight of the carrier gas that the ozone is being measured in. This value is used in the calculation of the wt% concentration unit. The default value is 32.0 (for Oxygen carrier gas.)
9) CONCENTRATION _RANGE
This variable sets the
10) TPC_ENABLE
This variable enables temperature and pressure compensation for the calculation of the ozone concentration. This variable should always be set to ON unless an external correction is being applied to compensate for temperature and pressure changes.
11)STD_TEMP
12)STD_PRESS
These variables are the standard temperature and pressure that are used to calculate the concentration when the units are set to grams per normal cubic meter (g/Nm3.)
13) RS232_MODE
This variable set the mode of operation of the RS232 interface. For more details on the use of the RS232 interface, please contact Teledyne API and request document number 01350.
14) CLOCK_ADJ
This variable is used to make slight adjustments to the instrument’s internal clock so that it keeps proper time. Setting this variable to a positive value will add that number of seconds each day to the clock while a negative value will subtract that number of seconds per day.
P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 37