Teledyne 450H Message/Test Measurement Field, Test Message, Test Measurements, Warning Message

Page 29
Message/Test Measurement Field

Message/Test Measurement Field

The message field shows warning messages or test measurements. Tables 2.2 and 2.3 summarize the test measurements and warning messages and their meanings. Refer to Section 2.3 for detailed information on viewing test measurements and warning messages and clearing warnings.

Test Message

 

Meaning

TIME=xx:xx:xx

 

Current time-of-day (HH:MM:SS)

RANGE = xxxx wt%

 

Full-Scale concentration for analog

 

 

output

O3 MEAS=xxxxx MV

 

Current UV reading, measure channel

O3 REF=xxxxx MV

 

Current UV reading, reference channel

PRESS=xxx PSIA

 

Absorption cell pressure - psia

GAS TEMP=xxx C

 

Temperature of the sample gas(deg. C)

LAMP TEMP=xxx C

 

UV Lamp Temperature (deg C)

BOX TEMP=xxx C

 

Internal box temperature (deg. C)

TABLE 2.2

TEST MEASUREMENTS

Warning Message

 

Meaning

 

SYSTEM RESET

 

Issued whenever Monitor is powered on

RAM INITIALIZED

 

RAM was erased

UV LAMP WARNING

 

UV lamp < 2500mV OR => 5000mV

UV LAMP SHUTDOWN

 

UV lamp temp control not working

SAMPLE FLOW WARNING

 

Sample flow < 700 cc/m

SAMPLE PRESSURE WARN

 

Sample pressure < 15 or > 35 In-Hg-A

GAS TEMP WARNING

 

Gas temp < 10 or > 50 deg. C

BOX TEMP WARNING

 

Box temp. < 12 deg. C or > 55 deg. C

ANA LAMP TEMP WARN

 

UV lamp temp < 51 or > 61 deg. C

V/F NOT INSTALLED

 

V/F card not installed or bad

TABLE 2.3

WARNING MESSAGES

Menu Field

The menu field changes depending on the mode of the Monitor and the buttons that have been pressed. It indicates the current function of each of the 8 pushbuttons below the display.

The 8 pushbuttons below the display are programmable by the CPU in that their functions change depending on the mode of the Monitor or the operations being performed. The legend above a button identifies its current function. If there is no legend above a button, it has no function and will be ignored if pressed.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 29

Image 29
Contents Printed 05/15/06 MODEL 450H OZONE MONITOREMAIL api-sales@teledyne.com SAFETY MESSAGES 3.0 SETUP MODE TABLE OF CONTENTS4.0 MAINTENANCE 5.0 ADJUSTMENTS5.4 CURRENT LOOP CALIBRATION APPENDIX A - ELECTRICAL SCHEMATIC INDEX5.3 A/D - D/A CALIBRATION PROCEDURE… FIGURES TABLE 1.1 DIGITAL OUTPUTS TABLESTABLE 1.2 CONTROL INPUTS TABLE 1.3 FINAL TEST AND CALIBRATION VALUES1.1Preface 1.0INTRODUCTIONADVANCED POLLUTION INSTRUMENTATION DIVISION 1.2 WARRANTY POLICY02024c COVERAGE1.3 Principle of Operation Page 1.5 Installation 19” Rack Mount Enclosure 1.4 SpecificationsPage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS 1.6Electrical I/O Connections FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS1.6.1Analog Output 1.6.2 Relay Outputs1.6.4 Control Inputs FIGURE 1.5 - CONNECTING DIGITAL OUTPUTSOutput DescriptionInput 1.6.5 RS232 OutputCONTROL INPUTS Zero CalibrateFIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUTU V L A M P M O T H E R B V / F C P U A C L I N E D C P O W E R SOPOWER SUPPLY MODULE Figure 1.8 - 19” Rack Assembly Layout1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8 Options 1.8.5 Flow Switch1.8.1 Auto-ZeroOption 1.8.2 Isolated Current Output 4-20mAor 2-20mA2.1.2 Concentration Alarms 2.1Key features2.1.4 E2 ROM backup of software configuration 2.0OPERATION2.2.1 Front panel display fields 2.2 Front Panel DisplayFIGURE 2.1 - MODEL 450H FRONT PANEL Mode FieldTest Message Message/Test Measurement FieldTEST MEASUREMENTS Warning Message2.2.2 Status LED’s 2.3 Software Operation2.3.1 Main Menu StateFIGURE 2.3 - ALARM STATUS MENU 2.3.3 Alarm Status MenuClearing Alarms Audible BeeperPage FIGURE 3.1 - SETUP MENU 3.0 SETUP MODE3.2 Setting the Concentration Units UNIT 3.1 Setting the Concentration Alarms ALRM3.3 Setting the time-of-dayand date CLK HI ALARM OFF3.4 Setting the RS-232baud rate COMM 0 LATCH_ALARMS 3.5 Setup variables VARS1 ALARM BEEPER 3 AZERO ENABLE10 TPC ENABLE 8 GAS MOLECULAR WEIGHT7 AZERO HOLD OFF 9 CONCENTRATION RANGE3.5.1Zero Calibration 3.5Diagnostics DIAG3.5.2 D/A Calibration 3.5.3 Analog Output4.2Cleaning Exterior Surfaces of the M450H 4.0Maintenance4.3 Cleaning the Optical Cell Filter Replacement Procedure4.4 Degree of Protection FIGURE 4.1 MEASUREMENT CELL DISASSEMBLYSwitch Settings 5.0 ADJUSTMENTS5.1 Calibration 5.2Changing the analog output rangeFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.3 A/D - D/A Calibration Procedure1.Press SETUP-DIAG 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page DESCRIPTION APPENDIX B - SPARE PARTS LISTPage Title APPENDIX C - ELECTRICAL SCHEMATIC INDEXDrawing Number