Teledyne 450H instruction manual

Page 12
1.Instruction manual.

1.Instruction manual.

2.Power Cord

CAUTION

To avoid personal injury, always use two persons to lift and

carry the Model 450H.

Upon receiving the Model 450H please do the following:

1.Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)

2.Before connecting power to the instrument, check the serial number tag to verify that the instrument has been configured for the correct line voltage and frequency connections.

CAUTION

Verify that the instrument is set up for proper line voltage and

frequency.

3.Connect an exhaust line to the fitting labeled ‘Exhaust.’ If the M450H has been configured with the ozone killer option, this line should be vented to an outside area, since the exhaust gas may still contain trace levels of ozone. If the M450H has not been equipped with an ozone killer, this line should be vented to an ozone destruct unit or other appropriate device.

CAUTION

Exhaust gas from the M450H may contain dangerous levels of

ozone!

4.Connect the ozone delivery line to the ¼” inlet fitting on the filter assembly on the rear panel. The ozone delivery pressure should be regulated to no more than 30psig. All tubing used should be made of ozone resistant material such as PTFE(Teflon) or PFA. If new, unconditioned tubing is used for the ozone delivery line, the ozone concentration may be affected until the tubing becomes conditioned. Pre-conditioned ¼” tubing can be purchased from Teledyne API. Contact the Teledyne API sales department and reference part number 02639.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 12

Image 12
Contents MODEL 450H OZONE MONITOR EMAIL api-sales@teledyne.comPrinted 05/15/06 SAFETY MESSAGES TABLE OF CONTENTS 3.0 SETUP MODE4.0 MAINTENANCE 5.0 ADJUSTMENTSAPPENDIX A - ELECTRICAL SCHEMATIC INDEX 5.3 A/D - D/A CALIBRATION PROCEDURE…5.4 CURRENT LOOP CALIBRATION FIGURES TABLES TABLE 1.1 DIGITAL OUTPUTSTABLE 1.2 CONTROL INPUTS TABLE 1.3 FINAL TEST AND CALIBRATION VALUES1.0INTRODUCTION 1.1Preface1.2 WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION02024c COVERAGE1.3 Principle of Operation Page 1.4 Specifications 1.5 Installation 19” Rack Mount EnclosurePage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS 1.6Electrical I/O Connections1.6.1Analog Output 1.6.2 Relay OutputsFIGURE 1.5 - CONNECTING DIGITAL OUTPUTS 1.6.4 Control InputsOutput Description1.6.5 RS232 Output InputCONTROL INPUTS Zero CalibrateFIGURE 1.6 - NEMA4X ASSEMBLY LAYOUT FIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAMU V L A M P M O T H E R B V / F C P U A C L I N E D C P O W E R SOFigure 1.8 - 19” Rack Assembly Layout POWER SUPPLY MODULE1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.5 Flow Switch 1.8 Options1.8.1 Auto-ZeroOption 1.8.2 Isolated Current Output 4-20mAor 2-20mA2.1Key features 2.1.2 Concentration Alarms2.1.4 E2 ROM backup of software configuration 2.0OPERATION2.2 Front Panel Display 2.2.1 Front panel display fieldsFIGURE 2.1 - MODEL 450H FRONT PANEL Mode FieldMessage/Test Measurement Field Test MessageTEST MEASUREMENTS Warning Message2.3 Software Operation 2.2.2 Status LED’s2.3.1 Main Menu State2.3.3 Alarm Status Menu FIGURE 2.3 - ALARM STATUS MENUClearing Alarms Audible BeeperPage 3.0 SETUP MODE FIGURE 3.1 - SETUP MENU3.1 Setting the Concentration Alarms ALRM 3.2 Setting the Concentration Units UNIT3.3 Setting the time-of-dayand date CLK HI ALARM OFF3.4 Setting the RS-232baud rate COMM 3.5 Setup variables VARS 0 LATCH_ALARMS1 ALARM BEEPER 3 AZERO ENABLE8 GAS MOLECULAR WEIGHT 10 TPC ENABLE7 AZERO HOLD OFF 9 CONCENTRATION RANGE3.5Diagnostics DIAG 3.5.1Zero Calibration3.5.2 D/A Calibration 3.5.3 Analog Output4.0Maintenance 4.2Cleaning Exterior Surfaces of the M450H4.3 Cleaning the Optical Cell Filter Replacement ProcedureFIGURE 4.1 MEASUREMENT CELL DISASSEMBLY 4.4 Degree of Protection5.0 ADJUSTMENTS Switch Settings5.1 Calibration 5.2Changing the analog output range5.3 A/D - D/A Calibration Procedure 1.Press SETUP-DIAGFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page APPENDIX B - SPARE PARTS LIST DESCRIPTIONPage APPENDIX C - ELECTRICAL SCHEMATIC INDEX Drawing NumberTitle