Teledyne 450H 1.6Electrical I/O Connections, 1.6.1Analog Output, Relay Outputs, Digital Outputs

Page 19
FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS

FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS

1.6Electrical I/O Connections

1.6.1Analog Output

The analog output is a two pin connector (See Figure 1.4). The analog output can be configured for voltage or current output. The standard is a 0-5 volt output. 0-100mV, 0- 1V, 0-10V, and 4-20 mA outputs are also available. See Section 5.2 for information on setting other output ranges.

1.6.2 Relay Outputs

Three form C relay outputs are provided on three pin connectors (See Figure 1.1). These outputs correspond to the ‘HI’ and ‘HI-HI’ concentration alarms and a System OK status output that is used to indicate a fault or error condition in the instrument. The relay contacts are rated to 3A at 240VAC. Do not exceed these ratings when connecting equipment to the instrument.

1.6.3 Digital Outputs

There are 12 digital output. These outputs are optically isolated NPN transistors which can pass 50 mA of DC current. These outputs can be used to interface to devices that accept logic-level digital inputs, such as Programmable Logic Controllers(PLC’s).

The outputs labeled ‘STATUS OUT’ are used to indicate instrument operational status and fault conditions. Table 1.1 below summarizes the functions of all the digital outputs.

P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 19

Image 19
Contents EMAIL api-sales@teledyne.com MODEL 450H OZONE MONITORPrinted 05/15/06 SAFETY MESSAGES 5.0 ADJUSTMENTS TABLE OF CONTENTS3.0 SETUP MODE 4.0 MAINTENANCE5.3 A/D - D/A CALIBRATION PROCEDURE… APPENDIX A - ELECTRICAL SCHEMATIC INDEX5.4 CURRENT LOOP CALIBRATION FIGURES TABLE 1.3 FINAL TEST AND CALIBRATION VALUES TABLESTABLE 1.1 DIGITAL OUTPUTS TABLE 1.2 CONTROL INPUTS1.1Preface 1.0INTRODUCTIONCOVERAGE 1.2 WARRANTY POLICYADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c1.3 Principle of Operation Page 1.5 Installation 19” Rack Mount Enclosure 1.4 SpecificationsPage 5.If the M450H has been configured with the auto-zerovalve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line 1.6 Installation NEMA 4X Wall Mount Enclosure ozone FIGURE 1.1 - NEMA 4X MOUNTING HOLE DIMENSIONS FIGURE 1.2 - NEMA 4X ELECTRICAL POWER CONNECTIONS FIGURE 1.3 - NEMA 4X PNEUMATIC CONNECTIONS 1.6.2 Relay Outputs FIGURE 1.4 - ELECTRICAL SIGNAL I/O CONNECTIONS1.6Electrical I/O Connections 1.6.1Analog OutputDescription FIGURE 1.5 - CONNECTING DIGITAL OUTPUTS1.6.4 Control Inputs OutputZero Calibrate 1.6.5 RS232 OutputInput CONTROL INPUTSD C P O W E R SO FIGURE 1.6 - NEMA4X ASSEMBLY LAYOUTFIGURE 1.7 - M450H PNEUMATIC BLOCK DIAGRAM U V L A M P M O T H E R B V / F C P U A C L I N EPOWER SUPPLY MODULE Figure 1.8 - 19” Rack Assembly Layout1.7 Operation verification TABLE 1.3 FINAL TEST AND CALIBRATION VALUES 1.8.2 Isolated Current Output 4-20mAor 2-20mA 1.8.5 Flow Switch1.8 Options 1.8.1 Auto-ZeroOption2.0OPERATION 2.1Key features2.1.2 Concentration Alarms 2.1.4 E2 ROM backup of software configurationMode Field 2.2 Front Panel Display2.2.1 Front panel display fields FIGURE 2.1 - MODEL 450H FRONT PANELWarning Message Message/Test Measurement FieldTest Message TEST MEASUREMENTSState 2.3 Software Operation2.2.2 Status LED’s 2.3.1 Main MenuAudible Beeper 2.3.3 Alarm Status MenuFIGURE 2.3 - ALARM STATUS MENU Clearing AlarmsPage FIGURE 3.1 - SETUP MENU 3.0 SETUP MODEHI ALARM OFF 3.1 Setting the Concentration Alarms ALRM3.2 Setting the Concentration Units UNIT 3.3 Setting the time-of-dayand date CLK3.4 Setting the RS-232baud rate COMM 3 AZERO ENABLE 3.5 Setup variables VARS0 LATCH_ALARMS 1 ALARM BEEPER9 CONCENTRATION RANGE 8 GAS MOLECULAR WEIGHT10 TPC ENABLE 7 AZERO HOLD OFF3.5.3 Analog Output 3.5Diagnostics DIAG3.5.1Zero Calibration 3.5.2 D/A CalibrationFilter Replacement Procedure 4.0Maintenance4.2Cleaning Exterior Surfaces of the M450H 4.3 Cleaning the Optical Cell4.4 Degree of Protection FIGURE 4.1 MEASUREMENT CELL DISASSEMBLY5.2Changing the analog output range 5.0 ADJUSTMENTSSwitch Settings 5.1 Calibration1.Press SETUP-DIAG 5.3 A/D - D/A Calibration ProcedureFIGURE 4.2- ANALOG OUTPUT JUMPERS 5.4 Current Loop Calibration zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter APPENDIX A - SOFTWARE MENU TREE Page DESCRIPTION APPENDIX B - SPARE PARTS LISTPage Drawing Number APPENDIX C - ELECTRICAL SCHEMATIC INDEXTitle