Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

AFGen2 To

This field has two subfields:

The upper field sets the destination port for audio frequency generator 2

FM -RF generator FM modulator

AM -RF generator AM modulator

Audio Out -AUDIO OUT connector

The lower field

sets FM modulation deviation if the upper field set to FM

sets AM modulation depth if the upper field set to AM

sets the amplitude of audio signal (volts RMS) at the AUDIO OUT connector if upper field is set to Audio Out

turns off the AF generator if the Yes On/Off key is pressed.

Operating Considerations

Simultaneous FM and AM, using both AF generators is not allowed.

AF generators 1 and 2 are automatically turned off any time the CDMA screens are displayed. This prevents accidental amplitude or frequency modulation of the CDMA carrier. The AF generators are turned back on when you exit the CDMA screens (if they were previously turned on).

Screens on which this field is displayed

RF GENERATOR

SIGNALING ENCODER (All Modes)

188

Chapter 6

C:\Spk\Ref\RefGuideRevE\Anlgscrn.fm

Page 184
Image 184