Operating Overview

Calibrating the Adjacent Channel Power (ACP) Measurement

Calibrating the Adjacent Channel Power (ACP) Measurement

To calibrate an ACP measurement using the internal CDMA generator:

Step 12. Remove power from the RF IN/OUT (or ANT IN) port.

Step 13. Go to the CDMA ANALYZER screen and choose the ACP measurement.

Step 14. Set the Tune Freq field (or RF Channel field) to the center frequency of your CDMA signal.

Step 15. Select the ACP Cal field and wait until the calibration is complete.

Chapter 3

63

Page 62
Image 62
Agilent Technologies 8935 series e6380a manual Calibrating the Adjacent Channel Power ACP Measurement