Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Control Fields for Analog Measurements

#Pts

This field sets the number of points to define or edit on the spectrum analyzer’s mask. You can define up to 15 points on the mask.

See “Using the Spectrum Analyzer Mask” on page 115 for more information about the spectrum analyzer mask.

Operating Considerations

The starting point of the mask is always the left graticule on the display, and the end point is always the right graticule on the display.

To define a straight-line mask, set the number of points to 1. The default mask is a straight line at the top of the display.

Screens on which this field is displayed

SPEC ANL: mask controls menu

AF Anl In

The audio frequency analyzer input selects the input for the analyzer. When selected, this field displays a list of choices.

Signals can be analyzed from three different types of inputs:

The output of the AM, FM, or SSB demodulators.

The AUDIO IN, ANALOG MODULATION IN, and AUDIO OUT connectors.

The signal present at the AM or FM modulators for the RF generator.

Operating Considerations

Changing this field changes the audio measurement (in the upper-right of the screen) to correspond to the selected signal.

See “Selecting the Oscilloscope’s Input” on page 120 for more information.

Screens on which this field is displayed

AF ANALYZER

SIGNALING DECODER (All Modes)