Overview of the Test Set’s Built-In Tools

Using the RF Tools Program

SA (Spectrum Analyzer) Self Calibration ON/OFF

This menu item allows you to check the calibration status of the Test Set’s spectrum analyzer, and change the status if you desire.

Figure 4-22 SA Self Calibration OFF/ON Screen

Select Main Menu

The RFTOOLS program uses the Test Set’s internal spectrum analyzer for making swept network analyzer measurements. The internal spectrum analyzer performs an automatic self calibration every 5 minutes. This automatic self calibration ensures that the spectrum analyzer operates within specified limits regardless of the operating temperature. However, this self calibration can introduce inaccuracy in spectrum analyzer output after the Test Set is operating at a constant temperature (after approximately 30 minutes).

Chapter 4

105