170 Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
RF Channel
The RF Channel field is displayed when the RF Display field on the
INSTRUMENT CONFIGURE screen is set to Chan.
This field displays the current channel number. Channel tuning
eliminates the need to enter transmit and receive frequencies directly
into the Test Set. Once the radio’s RF channel standard is selected, you
only have to enter the channel number to automatically set the RF
Generator and RF Analyzer to the correct frequency.
Operating Considerations
To utilize the correct transmit and receive frequencies select the
appropriate channel standard from the RF Chan Std field on the
INSTRUMENT CONFIGURE screen.
Screens on which this field is displayed
CDMA ANALYZER
CDMA GENERATOR
CODE DOM (Code Domain Analyzer): main controls menu
RF Gen Freq
This field specifies the center frequency of the generated CDMA signal.
This field is displayed only when the RF Display field on the
INSTRUMENT CONFIGURE screen is set to Freq.
NOTE Channel Tuning
You can configure the T est Set to allow direct ent ry of channel number s
(instead of entering the frequency) by using the RF Display field on the
INSTRUMENT CONFIGURE screen. See “RF Channel” on page 170
for more information.
Screens on which this field is present
CDMA GENERATOR
RF In/Ant
This field selects the Test Set’s RF input.
CAUTION The maximum safe level into the ANT IN connector is 60mW.
Exceeding this level could cause permanen t da m age to the Test Set.
Screens on which this field is present
CODE DOM (Code Domain Analyzer): main controls menu