CDMA Measurements - Screens and Control Fields

Control Fields for CDMA Measurements

Analyzer (CDMA analyzer)

The Analyzer field controls three functions:

Arm Meas arms the CDMA analyzer. The CDMA analyzer needs to be armed only when Single (see below) is selected.

Single/Cont selects between “single” measurements and “continuous” measurements.

When Single is chosen, you must arm each measurement by selecting the Arm Meas field (see above).

When Cont (continuous) is chosen, measurements are automatically armed.

The default selection is Cont.

Disarm disarms the CDMA analyzer during single operation. To stop continuous measurements, select Single. Selecting Disarm does not prevent the analyzer from being re-armed during continuous measurements

Screens on which this field is present

CDMA ANALYZER

Analyzer (code domain analyzer)

This field is used to arm or disarm measurements when making a single measurement (does not affect continuous measurements). Selecting Disarm stops the analyzer during a single measurement.

See “Measurement” on page 163 for more information.

Screens on which this field is present

COD DOM (Code Domain Analyzer): main controls menu

Anl Dir

This field is used to select the type of signal (forward or reverse link) to analyze.

Fwd sets the analyzer to look for base station transmitter data (forward link).

Rev sets the analyzer to look for mobile transmitter data (reverse link).

Default selection: Fwd

Screens on which this field is present

CDMA ANALYZER