Agilent Technologies 8935 series e6380a manual Defining the Fixed Mask’s Limits, Point

Models: 8935 series e6380a

1 320
Download 320 pages 25.58 Kb
Page 114
Image 114

Overview of theTest Set’s Built-In Tools

Using the Spectrum Analyzer

Defining the Fixed Mask’s Limits

1.Set the Edit Mask field to the limit you want to edit (Upper or

Lower).

You might want to set the Mask Beep field to Off while you are defining the mask.

2.Set the Mask Type field to Fix.

3.Set the #Pts field to the number of points you want defined by the mask. You can define up to 15 points. The starting point is always the left graticule on the display, and the end point is always the right graticule on the display.

To define a straight-line mask, set the number of points to 1. The default mask is a straight line at the top of the display.

4.Set the EditPt field to the point you want to define.

5.Set the Lvl and Freq fields to the frequency and level you want for the point you chose in the EditPoint field.

In Figure 4-27the frequency and level settings for each point are as follows:

Table 4-2

Settings for mask in Figure 4-27

 

 

 

 

 

 

Point

Level

Frequency

 

 

 

 

 

1

− 50 dBm

1799.7000

 

 

 

 

 

2

1 dBm

1799.7000

 

 

 

 

 

3

1 dBm

1800.3000

 

 

 

 

 

4

− 50 dBm

1800.3000 MHz

 

 

 

 

Figure 4-27 Setting the Spectrum Analyzer Mask

Point 2

Point 3

Point 4

Point 1

Set number of point here.

Select which point to set or edit.

Set the level of the

Set the frequency of the point.

 

Turn the beeper off or on.

 

Read the pass/fail

 

samask.pcx

 

 

116

Chapter 4

C:\Spk\Ref\RefGuideRevE\Tools.fm

Page 114
Image 114
Agilent Technologies 8935 series e6380a manual Defining the Fixed Mask’s Limits, Settings for mask in Figure, Point