Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Tune Freq

This field is displayed when the RF Display field on the INSTRUMENT CONFIGURE screen is set to Freq.

RF analyzer tune frequency sets the center frequency for the RF signal to be analyzed.

Screens on which this field is displayed

RF ANALYZER

TX Pwr Meas

The transmitter power measurement field specifies how transmitter power measurements are made:

Peak can be used to measure AM, FM, and un-modulated (CW) signals.

Sample can be used to measure FM or CW signals with increased measurement speed. This method can only be used with signals having no amplitude modulation component.

Operating Considerations

If you change this field, zero the power measurement (using the TX Pwr Zero field) before measuring power.

Screens on which this field is displayed

RF ANALYZER