Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Auto Zero

This field is displayed for average power measurements.

Auto is used to automatically zero the TX power measurement periodically during operation.

Manual is used to override the Auto feature. When Manual is selected, you must manually zero the TX power measurement by selecting the TX Pwr Zero field.

Screens on which this field is present

RF ANALYZER

BW=

This field displays the spectrum analyzer’s resolution bandwidth. The resolution bandwidth and sweep rate are determined by the span setting, and cannot be set independently. See “Setting Resolution Bandwidth and Sweep Rate” on page 111 for more information.

Screens on which this field is displayed

SPEC ANL: all menus except mask controls

Center Freq

This field sets the tune frequency for the center of the spectrum analyzer’s screen. The center frequency field is displayed when the RF Display field on the INSTRUMENT CONFIGURE screen is set to Freq.

Operating Considerations

This field also changes the center frequency of the tracking generator and the RF analyzer.

Screens on which this field is displayed

SPEC ANL: main controls menu

Chapter 6

191

Page 187
Image 187
Agilent Technologies 8935 series e6380a Center Freq, Spec ANL all menus except mask controls, Spec ANL main controls menu