CDMA Measurements - Screens and Control Fields

Control Fields for CDMA Measurements

Control Fields for CDMA Measurements

# of Frames

See “Data Source” on page 157.

ACP Cal

This field calibrates the ACP measurement for the current RF analyzer frequency at the selected frequency offset (ACP Offset) and filter bandwidth (ACP Fltr BW).

This field is displayed only when the ACP measurement is selected.

You must always remove power at the ANT IN or RF IN/OUT connector before calibrating adjacent channel power to ensure accurate calibration.

Operating Considerations

The Test Set can memorize approximately 16 sets of ACP calibration data. If calibration data does not exist for a given channel and offset setting, the following message will be displayed:

ACP UNCAL: No Freq or ACP Offset cal. Start ACP Cal.

To clear this message, select the ACP Cal field to start the calibration. When the calibration is finished, the message will be gone.

Screens on which this field is present

CDMA ANALYZER

ACP Fltr BW

This field designates the filter bandwidth around the adjacent channels at which the measurements are made. Valid values are 10 kHz to 1.23 MHz with 100 Hz resolution.

This field is displayed only when the ACP measurement is selected.

Operating Considerations

Power is only measured with bandwidths up to 100 kHz. If a wider bandwidth is specified, the power in the first 100 kHz is measured and used to calculate the power in the total specified bandwidth.

Screens on which this field is present

CDMA ANALYZER