Configuration - Screens and Control Fields

Control Fields on the Configuration Screens

Control Fields on the Configuration Screens

Abort Print

Select this field to interrupt the printing job in progress.

Abort Print is not directly programmable over the GPIB.

Screen(s) Where Field is Present

PRINTER CONFIGURE

Antenna In

This field is used to indicate losses or gains between the ANT IN port and the device-under-test.

Enter a positive value to indicate a gain (such as an amplifier). The spectrum analyzer’s marker level (Lvl) measurement is automatically reduced by that amount. The spectrum analyzer’s Ref Level is automatically decreased by the same amount, so the trace position does not appear to change.

Enter a negative value to indicate a loss (such as cable loss). The spectrum analyzer marker’s level (Lvl) measurement is automatically increased by that amount. The spectrum analyzer’s Ref Level is automatically increased by the same amount, so the trace position does not appear to change.

This field is only used when the RF Level Offset field is set to On. See “RF Level Offset” on page 246.

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE

Chapter 7

227

Page 223
Image 223
Agilent Technologies 8935 series e6380a manual Control Fields on the Configuration Screens, Abort Print, Antenna