Connectors

Connectors

 

RF IN/OUT

 

This Type-N connection is used to output signals from the RF generator and to input RF

 

signals.

 

 

CAUTION

Overpower Damage — Refer to the Test Set’s connector panel for maximum input power

 

level. Exceeding this level can cause permanent instrument damage.

 

If the RF power at the RF IN/OUT port exceeds allowable limits, a loud warning signal

 

sounds and a message appears at the top of the screen. If this occurs, disconnect the RF

 

power, reset using Meas Reset (press and release the Shift key, then the Hold key), and

 

allow the Test Set to cool off for approximately 2 minutes before making any other

 

measurements on this port.

 

Operating Considerations

 

This port must be used when measuring TX (RF) power.

 

Signals ≤ 60 mW can be input to the ANT IN connector for all RF measurements except

 

TX Power.

 

This port can be selected on the RF ANALYZER, RF GENERATOR, CDMA

 

GENERATOR, CDMA ANALYZER, CODE DOM, or SPEC ANL screens.

 

SCOPE MONITOR OUT

 

This connector provides an external output from the AF analyzer.

 

Operating Considerations

 

The Scope To field in the AF ANALYZER screen determines the source of this signal.

 

For more information see the “Scope To” on page 214.

 

The level is not affected by the front-panel VOLUME knob.

 

Output impedance <1 kΩ