Overview of theTest Set’s Built-In Tools

Using the RF Tools Program

Each RFTOOLS test includes two parts: calibration and measurement. If the spectrum analyzer performs a self calibration between the calibration and measurement of the RFTOOLS test, the spectrum analyzer’s automatic self calibration instability will increase RFTOOLS test measurement error. To avoid this problem, you may use this menu item to turn off the spectrum analyzer’s self calibration. This self calibration should only be turned off after the unit has reached a constant powered-on operating temperature (approximately 30 minutes after power-on), therefore it is probably best to leave it on. After 30 minutes, turning the self calibration off will improve the measurement accuracy of RFTOOLS tests.

Catalog PC (Memory) Card

This utility lists all the files contained on the PC card. The PC card that you want to catalog must be inserted prior to selecting this test.

File types are denoted by the following suffixes:

.PGM suffix denotes a program code file

.LIB suffix denotes a library file

.PRC suffix denotes a procedure file

.DAT suffix denotes a data file

.PLT suffix denotes a plot file

106

Chapter 4

C:\Spk\Ref\RefGuideRevE\Tools.fm

Page 104
Image 104
Agilent Technologies 8935 series e6380a manual Catalog PC Memory Card