CDMA Measurements - Screens and Control Fields

Control Fields for CDMA Measurements

Chan Size

This field appears only in IS-2000 Code Domain measurements. The size reflects its location in the Walsh set, with 1 indicating that there is only one channel included. By comparison, a channel size of 32, will encompass (and therefore make unavailable) a total of 32, channels.

Screens on which this field is present

COD DOM (Code Domain Analyzer, IS-2000 only)

CD pwr unit

This field allows you to select absolute or relative power measurements for the code domain analyzer. See “Code Domain Power, IS-95” on page 139 for more information about these measurements.

Screens on which this field is present

CODE DOM (Code Domain Analyzer)

Chn Pwr Cal

This field is displayed when the channel power measurements are selected. It is used to calibrate the channel power measurement. When this field is selected, an internally generated calibration signal is measured using the average power technique. Channel power is also measured and a correction factor is generated. This correction factor is applied to subsequent channel power measurements.

Calibration should be performed whenever a new set of measurements is made and whenever the frequency of the measured signal is changed. See “Calibrating the Channel Power Measurement – CDMA Analyzer” on page 61 for more information on analyzer calibration and “Calibrating the Channel Power Measurement – Code Domain” on page 62 for code domain calibration.

Screens on which this field is present

CDMA ANALYZER

CODE DOM (Code Domain Analyzer, IS-95 option): CD Setup Control menu, Abs selected for CD pwr unit.

CODE DOM (Code Domain Analyzer, IS-2000 option): Reference Control menu, Abs selected for CD pwr unit.