Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Amplitude

This field adjusts the amplitude of the RF generator.

See for Atten Hold (below) for more information.

Screens on which this field is displayed

RF GENERATOR

SPEC ANL: RF generator controls menu

Atten Hold

Attenuator hold prevents the fixed RF output attenuators from switching in and out, eliminating the loss of the output signal as the level is changed. This function is helpful when making squelch measurements.

Operating Considerations

When this function is set to On, the RF output level is restricted to a range around the present Amplitude setting. This range varies with the amplitude setting. Attempting to set an amplitude outside the allowed range results in an error message and beep (if the beeper is on). RF output level accuracy is greatly degraded outside the allowed range.

Screens on which this field is displayed

RF GENERATOR