Connectors

Connectors

 

DATA IN

 

This connector provides a data input to the CDMA generator.

 

Operating Considerations

 

Input level = TTL

 

Input impedance = 50k Ω

 

Data Rate = 1.2288 Mcps

 

DUPLEX OUT

 

This connector is an output for the RF generator, CDMA generator and tracking

 

generator.

 

Operating Considerations

 

Output impedance = 50 Ω

 

The RF generator’s output is selected in the Output Port field. This field is available on

 

the RF GENERATOR, CDMA GENERATOR, and SPEC ANL screens.

 

 

CAUTION

Connecting a signal of >60 mW (17.78 dBm) to the DUPLEX OUT port can cause

 

instrument damage (although internal protection circuits can typically withstand a

 

short-duration signal of several Watts).

 

If the overpower circuit is triggered (signified by a warning message at the top of the

 

screen), remove the signal from the DUPLEX OUT port, and reset using Meas Reset

 

(press and release the Shift key, then the Hold key) or turn the Test Set off and on to reset

 

it.

 

EVEN SECOND SYNC IN

 

This port is an input for the even second clock from the base station under test. A

 

positive edge on this connector starts two timers. One timer, after reaching terminal

 

counts, starts CDMA frame clocks. The other timer starts pilot PN sequence generation.

 

The Even Sec In field on the CDMA GENERATOR screen controls this port. See “Even

 

Sec In” on page 158.

 

Operating Considerations

 

Input impedance = 50 kΩ

 

Input level range = TTL

286

Chapter 9

Page 282
Image 282