CDMA Measurements - Screens and Control Fields

Control Fields for CDMA Measurements

Even Sec In

This field controls the internal connection of the EVEN SECOND SYNC IN connector to the CDMA analyzer or generator.

When Enable is selected, signals applied to the EVEN SECOND SYNC IN connector synchronize the timing circuits.

When Not is selected, input signals applied to the EVEN SECOND SYNC IN are ignored. This mode may be useful when signals other than an even-second clock are used for synchronization. For example, when using a one-second clock, you would select Not after synchronization to prevent the internal even-second clock from being retriggered half-way through its cycle.

Screens on which this field is present

CDMA ANALYZER

CDMA GENERATOR

Find PN

This field selects whether the PN offset will be automatically calculated (Auto), or manually entered (Manual). (See “PN Incrment” on page 165 and “PN Offset” on page 165 for more information.)

Screens on which this field is present

CODE DOM (Code Domain Analyzer)

Freq Err

This measurement display field displays the frequency error. See “Frequency Error (Freq Err)” on page 134 for more information about this measurement.

Screens on which this field is present

CODE DOM (Code Domain Analyzer)

CDMA ANALYZER

CDMA GENERATOR