Agilent Technologies 8935 series e6380a Cdma Measurements, Average Power Measurements, Avg Pwr

Models: 8935 series e6380a

1 320
Download 320 pages 25.58 Kb
Page 128
Image 128

CDMA Measurements - Screens and Control Fields

CDMA Measurements

CDMA Measurements

Average Power Measurements

Avg Pwr

Avg Pwr displays the average power level of all signals at the RF IN/OUT port. The units (watts, volts, or dBm) can be changed by positioning the cursor in front of the units field and then pressing the desired units key. (See “Input Port” on page 201.) Refer to the Agilent Technologies 8935 CDMA Base Station Tests Applications Guide for more information about average power measurements.

Screens on which this measurement is displayed

CDMA ANALYZER

CDMA GENERATOR

ADC FS

ADC FS is displayed when average power is measured. ADC FS indicates how close the measured power level is to the maximum allowed input level at the RF IN/OUT port for a particular Pwr Gain setting (on the CDMA ANALYZER screen). Changes to ADC FS can be made by altering the Pwr Gain field setting and by altering the signal level into the RF IN/OUT port. ADC FS should be between − 1.0 and −10 dB when the Pwr Gain field is set to Auto, and can be manually set up to − 0.5 dB when set to Hold. The measurement will be aborted if ADC FS goes above − 0.5 dB. Measurement accuracy may be degraded if ADC FS goes below − 10 dB.

Screens on which ADC FS is displayed in conjunctions with Average Power measurements

CDMA ANALYZER

CDMA GENERATOR