Operating Overview

Saving and Recalling Instrument Setups

To remove (clear) an individual save register

Step 1. Specify where the register is stored using the Save/Recall field on the I/O CONFIGURE screen.

Step 2. Press Recall.

Step 3. Use the knob to position the cursor in front of the register to be removed from the Recall menu at the bottom-right of the screen. The register name and percentage of memory occupied by that register are indicated at the top of the screen.

Step 4. Press Yes On/Off. A prompt appears, asking if you want to delete the save register.

Step 5. Press Yes On/Off.

To clear all save registers

Step 1. Press Recall.

Step 2. Use the knob to position the cursor in front of the *Clr All* entry in the Recall menu at the bottom-right of the screen.

Step 3. Press the knob or press Enter. A prompt appears at the top of the screen to verify that you want to clear all registers.

Step 4. Press Yes On/Off.

Register Names

You can use any number, letter, or combination of numbers and letters as a name for storing instrument settings. For instance; if you want to save a setup for testing a “Vulcan7” radio, you can save the setting as “VULCAN7”.

Two register names are reserved for special purposes: POWERON and BASE.

POWERON Settings

When the Test Set is turned on, it uses a set of instrument setup parameters specified at the time of manufacture. You can have the instrument power up in a different state by making the desired changes to the original settings, and then saving them using the name

POWERON.

The next time the instrument is turned on, the instrument returns to the state present when you saved the POWERON setting. For instance; if the SCOPE screen was displayed when POWERON was saved, it is the screen that is displayed when you turn the instrument on.