Agilent Technologies 8935 series e6380a manual Message FVC NAMPS-TACS, Message/DST FVC

Models: 8935 series e6380a

1 320
Download 320 pages 25.58 Kb
Page 266
Image 266

Signaling Encoder and Decoder - Screens and Control Fields

Control Fields for the Signaling Encoder and Decoder

Message (FVC) (NAMPS-TACS)

Figure 8-8 NAMPS-NTACS FVC Message Data output Format

DSAT 24-bits

Send

Sync word 30-bits

FVC

MessageDSAT

12345672556B

Message Info

 

Parity

 

 

28--bits

 

12-bits

 

 

 

 

 

DSAT DSATDSAT

24-bits

24-bits

24-bits

 

 

 

Data Direction

This description applies to the NAMP-NTAC mode.

The 7 hexadecimal characters (28 bits) of this FVC message are combined with 12 parity bits calculated by the encoder to output a 40-character data stream. This information is output when the Message/DST field is set to Message, and Send is selected.

Operating Considerations

The entire field must contain data, no blank spaces are allowed. DSAT is turned off while the FVC Message Stream is being sent.

Screens Where Field is Present

SIGNALING ENCODER (Modes: NAMP-NTAC)

Message/DST (FVC)

This field is available only in NAMP-NTAC mode.

This field determines what type of data is sent when Send is selected:

1.Selecting Message causes the contents of the Message field to be output.

2.Selecting DST causes the digital signaling tone sequence to be output. The sequence sent is the inverse of the sequence entered in the DSAT field, and is automatically determined by the Encoder.

The DST values are: DAA934, DAA4D4, DA9564, DA52B2, D954D4, D94D52, and D69654.

Screens Where Field is Present

SIGNALING ENCODER (Modes: NAMP-NTAC)

270

Chapter 8

C:\Spk\Ref\RefGuideRevE\Enc-dec.fm

Page 266
Image 266
Agilent Technologies 8935 series e6380a manual Message FVC NAMPS-TACS, Message/DST FVC