CDMA Measurements - Screens and Control Fields

CDMA Measurements

Screens on which this measurement is displayed

• CODE DOM (Code Domain Analyzer, IS-2000)

Fast Power

Fast power is a faster method of measuring code domain power. A number of internal values must be transferred from a non-fast power measurement before fast measurements begin, or when changes to any of the following fields occur: Data Rate, PN Offset, CDMA TB, Synth Ref.

The Fast Power Synchronize measurement is designed to set these internal variables.

The number of measurement averages can be entered in the Num Avgs field on the FP Setup menu. This number is used to calculate and display successive code domain power measurements.

Screens on which this measurement is displayed

• CODE DOM (Code Domain Analyzer)

Code Domain Power & Noise

The Power & Noise option displays the Code Domain channels along with channel noise, shown as hatch marks within each channel area. Noise and power are combined in supplemental channels to take into account the relationship between channel width and noise power. For example, a supplemental channel with a width of 2, in which each individual channel has a noise power value of -15 dBm, will have a combined noise power value of -12 dBm, or an increase of roughly 3 dB.

Figure 5-14 Code Domain Power & Noise

Screens on which this measurement is displayed

• CODE DOM (Code Domain Analyzer)