Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Squelch

This setting determines the squelch operation when demodulating FM, AM, or SSB signals. Three settings are available:

Pot uses the front-panel SQUELCH knob for squelch level adjustment.

Open disables squelch operation.

Fixed sets the squelch to a fixed level, disabling the front-panel SQUELCH knob control.

Operating Considerations

Most measurements are not displayed on the screen if the incoming signal falls below the squelch level (with the exception of TX Power). The measurements are replaced by four dashes (- - - -) to indicate they have been squelched.

Oscilloscope measurements are also disabled when the signal has been squelched.

Spectrum analyzer measurements are not affected by the squelch setting (although squelch still affects whether the demodulated signal can be heard while viewing the RF signal).

Trying to read a squelched measurement using GPIB will cause your program to halt until the squelch is either turned down, a measurement is made, or until a program time-out aborts the measurement.

Screens on which this field is displayed

RF ANALYZER

Chapter 6

217