Overview of the Test Set’s Built-In Tools

Using the RF Tools Program

Swept Insertion Loss

Measurement Description

This test measures the loss of a cable or device by sweeping it over the specified frequency range. This test uses the Spectrum Analyzer to make measurements.

Making the Measurement

1.Select Swept Insertion Loss from the Select Test (Main) Menu. A Swept Insertion Loss Information menu will appear.

Figure 4-10 Selecting the Swept Insertion Loss Measurement

Select Test Menu

Swept Insertion Loss

2.Position the cursor at Start frequency in MHz and select it. Enter a value for the start of the frequency sweep from 0.4 MHz to

1000 MHz or 1700 MHz to 2000 MHz.

3.Position the cursor at Stop frequency in MHz and select it. Enter a value from 0.4 MHz to 1000 MHz or 1700 MHz to 2000 MHz. This value should be larger than the start frequency.

Chapter 4

89

Page 87
Image 87
Agilent Technologies 8935 series e6380a manual Selecting the Swept Insertion Loss Measurement