Operating Overview

Using Channel Numbers to Set Analyzer and Generator Frequencies

Using Channel Numbers to Set Analyzer and Generator Frequencies

RF analyzer and RF generator frequencies can be entered by channel number (the default setting) or by discrete frequencies (in MHz). The RF Display field on the INSTRUMENT CONFIGURE screen controls which way frequencies are entered. This screen is accessed by pressing the Inst Config key.

If the RF Display field is set to Chan for channel tuning, you also need to set the RF Chan Std field to indicate the standard for your base station (such as North American PCS). This automatically sets the correct frequency offset needed to generate reverse channel signals and analyze forward channel signals for base stations.

If you do not know the channel number of your base station, but know the transmit and receive frequencies, set the RF Display field to Freq. You can then enter these frequencies directly for the RF generator and RF analyzer during tests.

Figure 3-9 Configuration to Use Channel Numbers for RF Generator and Analyzer Settings

Set the RF Display field to Chan to turn on channel tuning.

Select the RF Chan Std field to

display a list of system types.....

..and then select the channel standard for testing your base stations.

60

Chapter 3

C:\Spk\Ref\RefGuideRevE\Opoverv.fm

Page 59
Image 59
Agilent Technologies 8935 series e6380a manual Operating Overview