Analog Measurements - Screens and Control Fields

Control Fields for Analog Measurements

Reset

Reset retriggers the measurement when Single triggering is selected. See “Cont/Single” on page 192.

Screens on which this field is displayed

SCOPE: trigger controls menu

RF Channel

The RF Channel field is displayed on the screens listed below when the RF Display field on the INSTRUMENT CONFIGURE screen is set to Chan.

This field displays the current channel number. Channel tuning eliminates the need to set transmit and receive frequencies directly into the Test Set. Once the radio’s RF channel standard is selected, you only have to enter the channel number to automatically set the RF Generator and RF Analyzer to the correct frequency.

Operating Considerations

To utilize the correct transmit and receive frequencies select the appropriate channel standard from the RF Chan Std field on the INSTRUMENT CONFIGURE screen.

The lower subfield of the RF Channel field on the spectrum analyzer’s RF Gen controls menu also sets the RF channel standard.

Screens on which this field is displayed

RF ANALYZER

SPEC ANL: RF generator or main controls menu

RF Cnt Gate

RF counter gate specifies how long the RF counter samples the signal before displaying the frequency. Specifying a shorter gate time may enable you to see frequency fluctuations that might not be seen using a longer gate time.

Screens on which this field is displayed

RF ANALYZER

212

Chapter 6

C:\Spk\Ref\RefGuideRevE\Anlgscrn.fm

Page 208
Image 208
Agilent Technologies 8935 series e6380a manual Reset, RF Cnt Gate, Spec ANL RF generator or main controls menu