Configuration - Screens and Control Fields

Control Fields on the Configuration Screens

RF Offset

This field is displayed when the RF Display field is set to Freq.

This field enables/disables the RF generator − RF analyzer frequency offset specified in the (Gen)-(Anl)field below it.

Operating Considerations

When an RF offset is used, changing the RF generator’s frequency or RF analyzer’s tune frequency automatically alters the other setting.

See “Setting an RF Generator/Analyzer Offset” on page 69 for more information.

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE

Save/Recall

This field specifies which memory device the Test Set accesses when the save and recall functions are used.

Internal is a section of internal RAM. RAM is also used for running IBASIC programs, which may require you to delete the save/recall registers if the program is very large.

Card is the front-panel MEMORY CARD (PC card) slot. A

Save/Recall Device is not Present message is displayed if you try to save or recall an instrument setup when a write-able PC card is not installed in the Test Set.

RAM refers to RAM disks that you can create on internal RAM. Refer to the Programmer’s Guide for information on creating RAM disks. This is part of the same memory used when “internal” is specified, and may have to be erased when loading very large IBASIC programs. A Save/Recall Device is not initialized message is displayed if you try to save or recall an instrument setup when a RAM disk has not been created.

Screen(s) Where Field is Present

I/O CONFIGURE

Chapter 7

247

Page 243
Image 243
Agilent Technologies 8935 series e6380a manual RF Offset, Save/Recall