CDMA Measurements - Screens and Control Fields

Control Fields for CDMA Measurements

 

PN Offset or PN Ofs

 

This measurement display field (PN Offset) is displayed on the CDMA

 

ANALYZER and CDMA GENERATOR screens when Rho

 

measurements are selected. If the PN Offset measurement is not

 

displayed, select the Freq Err measurement field, the choose PN Offset.

 

The same measurement is displayed in the PN Ofs measurement field

 

on the CODE DOM screen.

 

See “Find PN” on page 158 for more information.

 

When the Even Sec In field is set to Not, changes to the PN Offset

NOTE

 

field have no effect on the analyzer.

 

Screens on which this field is present

 

 

CDMA ANALYZER

 

CDMA GENERATOR

 

CODE DOM (Code Domain Analyzer): PN setup controls menu

 

Pwr Gain

 

This field is used to set the level into the CDMA analyzer when making

 

average power measurements. Levels are set in 6 dB increments. The

 

IF signal level is displayed as ADC FS when average power is measured,

 

and should be between -1.0 and- 5 dB for best measurement results.

 

Auto automatically sets the gain for an ADC FS value in the range of

 

− 1 to − 5 dB. This is the recommended mode of operation.

 

Hold lets you override the automatic setting to manually set the

 

gain.

 

This field is only displayed when an average power measurement is

 

selected.

 

Screens on which this field is present

 

CDMA ANALYZER

166

Chapter 5

C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm

Page 163
Image 163