Overview of theTest Set’s Built-In Tools

Using the RF Tools Program

3.Position the cursor at DUPLEX OUT Level in dBm and enter the power level at which the test should be conducted.

4.Press k1 (Begin Tst) to begin the test.

Figure 4-9 Discrete Frequency Insertion Loss Setups

Calibration Setup

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

 

 

HI

 

LO

 

 

 

 

 

 

I

Q

IN

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

IN

OUT

 

 

 

 

 

8935

ANT IN

DUPLEX OUT

16X

 

FRAME

EVEN

TRIGGER

 

 

 

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

Measurement Setup

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

 

 

HI

 

LO

 

 

 

 

 

 

I

Q

IN

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

IN

OUT

 

 

 

 

 

8935

ANT IN

DUPLEX OUT

16X

 

FRAME

EVEN

TRIGGER

 

 

 

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

ANT IN

6dB Pad

Cal Cable

sweptg.eps

DUPLEX OUT

6dB Pad

ANT IN

6dB Pad

Cal Cable

DUPLEX OUT

6dB Pad

Test Cable

or Device

Barrel

Connector

5. A screen appears with a setup diagram. (Figure 4-9) This test requires two 6-dB pads. The pads are put on the ANT IN and DUPLEX OUT ports on the Test Set to reduce impedance mismatch error at these ports. A reference level is obtained by connecting a short calibration cable between the pads on the DUPLEX OUT and ANT IN ports. Press k1 (Proceed).

 

6.

After the Test Set has finished calibrating, a second diagram

 

 

appears on the screen. Connect the test cable or device to be tested

 

 

between the calibration cable and one of the pads. Press k1

 

 

(Proceed).

 

7.

The discrete frequency insertion loss results screen appears with the

 

 

results of the test. The loss is expressed as a negative number in this

 

 

particular test.

 

The insertion loss at the specified frequency is displayed twice. The

NOTE

 

second display is enlarged for easier reading.

 

8.

Press k1 (Proceed) when you are finished with this screen. The

 

 

Discrete Frequency Insertion Loss Test menu appears. If you would

like to repeat the measurement press k1 (Begin Tst). The calibration step will be skipped if none of the parameters were changed. To exit the test press k5 (Return).

88

Chapter 4

C:\Spk\Ref\RefGuideRevE\Tools.fm

Page 86
Image 86