Agilent Technologies 8935 series e6380a manual Level value to avoid damage to that device

Models: 8935 series e6380a

1 320
Download 320 pages 25.58 Kb
Page 83
Image 83

Overview of the Test Set’s Built-In Tools

 

 

Using the RF Tools Program

 

6.

Position the cursor at DUPLEX OUT Level in dBm and select it.

 

 

Enter the power level, − 54 to +10 dBm, at which the test should be

 

 

conducted. If you are testing a power sensitive device, enter a low

 

 

level value to avoid damage to that device.

 

Damage may result if the power input to the Test Set’s ANT IN port

CAUTION

 

exceeds 60 mW (which is equivalent to 18 dBm). Damage may also

 

result to the unit-under-test if it is overdriven by the DUPLEX OUT

 

power level. To avoid damage to the unit under test, enter a low power

 

level for the DUPLEX OUT level in dBm parameter.

 

7.

Press k1 (Begin Tst).

Figure 4-5 Swept Gain Setups

Calibration Setup

Measurement Setup

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

P A R A L L E L P O R T 1 5

S E R I A L 9

 

 

HI

 

LO

 

 

 

 

 

 

 

 

 

 

 

P A R A L L E L P O R T 1 6

S E R I A L 1 0

 

 

 

 

I

Q

IN

 

 

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

 

S E R I A L 1 1

 

 

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

 

 

IN

OUT

 

 

 

 

 

 

 

8935

 

 

16X

 

FRAME

EVEN

TRIGGER

 

 

 

ANT IN

DUPLEX OUT

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

P A R A L L E L P O R T 1 5

S E R I A L 9

 

 

HI

 

LO

 

 

 

 

 

 

 

 

 

 

 

P A R A L L E L P O R T 1 6

S E R I A L 1 0

 

 

 

 

I

Q

IN

 

 

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

 

S E R I A L 1 1

 

 

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

 

 

IN

OUT

 

 

 

 

 

 

 

8935

 

 

16X

 

FRAME

EVEN

TRIGGER

 

 

 

ANT IN

DUPLEX OUT

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

ANT IN

6dB Pad

Cal Cable

sweptg.eps

DUPLEX OUT

6dB Pad

ANT IN

6dB Pad

Cal Cable

DUPLEX OUT

6dB Pad

Test Device

Barrel

Connector

8.A setup screen appears with a diagram of the necessary connections, similar to the Calibration Setup of Figure 1-2. This test requires two 6-dB pads (attenuators). The pads are put on the Test Set’s ANT IN and DUPLEX OUT ports to reduce impedance mismatch error at these ports. A reference level is obtained by connecting a short calibration cable between the pads on the DUPLEX OUT and ANT IN ports. Set up the hardware, then press k1 (Proceed) or the knob to continue.

9.Another setup screen appears similar to the Measurement Setup of Figure 4-5. Attach the device to be tested between the calibration cable and one of the pads. In this step, a measurement is made of the additional gain from the reference level. Press k1 (Proceed) or the knob to continue.

Chapter 4

85

Page 83
Image 83
Agilent Technologies 8935 series e6380a manual Level value to avoid damage to that device, Press k1 Begin Tst