Operating Overview

Calibrating the Channel Power Measurement – Code Domain

Calibrating the Channel Power Measurement

– Code Domain

To calibrate a channel power measurement using the internal CDMA generator:

Step 8. Remove power from the RF IN/OUT (or ANT IN) port.

Step 9. Go to the CODE DOMAIN screen and choose the CD Setup (IS-95 only) or Reference (IS-2000) control.

Step 10. Select Abs from the CD pwr unit field.

Step 11. Select the Chn Pwr Cal field and and press the knob to start calibration. Wait until the calibration is complete.