Signaling Encoder and Decoder - Screens and Control Fields

Control Fields for the Signaling Encoder and Decoder

Message (FVC) (AMPS-TACS)

Figure 8-7 AMPS-TACS FVC Message Data Output Format

The data you enter here............

F V C

Message

T1 T2 (2 Bits)

Information

(26 Bits)

Parity

(12 Bits)

Contents of the Message or Field

 

Message Word

.........is output in this sequence

 

101 Bits

11 Bits 40 Bits

37 Bits 11 Bits 40 Bits

DOTTING

W.S.

REPEAT 1

OF

WORD

DOTTING W.S.

REPEAT 2

OF

WORD

W.S.=Word Sync

37 Bits

11 Bits

40 Bits

37 Bits 11 Bits

40 Bits

37 Bits

11 Bits

40 Bits

DOTTING W.S.

REPEAT 9

OF

WORD

D O T T IN G

W.S.

REPEAT 10

OF

WORD

DOTTING W.S.

REPEAT 11

OF

WORD

This description applies to the AMPS-TACS mode.

The 7 hexadecimal characters of the FVC Message field represent the 2 type bits and 26 information bits in the FVC message word.

The generation of dotting, word sync, parity, and the 11 repetitions of these parameters in the FVC message stream is done automatically.

Operating Considerations

The entire field must contain data, no blank spaces are allowed. SAT is turned off while the FVC message stream is being sent.

Screens Where Field is Present

SIGNALING ENCODER (Modes: AMPS-TACS)

Chapter 8

269

Page 265
Image 265
Agilent Technologies 8935 series e6380a manual Message FVC AMPS-TACS, Signaling Encoder Modes AMPS-TACS