Overview of theTest Set’s Built-In Tools

Using the RF Tools Program

10.The trace of the device gain over the frequency range swept is displayed. If the gain exceeds 40 dB, it will not appear on the trace. The maximum gain is also displayed as text on the screen. To save the plot to a PC card, press k4 (Save Plot) and use the knob to enter a filename. Press k5 (Scr Dump) to send the image to a printer or to the BTS Laptop Utility. Directions will appear at the top of the screen. Press PRINT and allow the image to print to the connected device. Then, press PAUSE/CONTINUE. Press k1(Proceed) when you are ready to continue.

Figure 4-6 Example of a Swept Gain Plot Screen

11.The swept gain results screen (see Figure 4-7) appears with the test results. If the BTS Laptop Utility is in use, this data is collected by the laptop. If you wish to repeat the test using the same calibrated devices and test parameters, press k1 (Repeat). To re-display the plot screen, press k3 (Disp Plot). To return to the Select Test Menu, press k5 (Return).

Figure 4-7 Example of a Swept Gain Results Screen

86

Chapter 4

C:\Spk\Ref\RefGuideRevE\Tools.fm

Page 84
Image 84